Analysis of temperature dependence of background phase errors in phase-contrast cardiovascular magnetic resonance


Busch, Julia; Vannesjo, Johanna S; Barmet, Christoph; Pruessmann, Klaas P; Kozerke, Sebastian (2014). Analysis of temperature dependence of background phase errors in phase-contrast cardiovascular magnetic resonance. Journal of Cardiovascular Magnetic Resonance, 16:97.

Abstract

Background: The accuracy of phase-contrast cardiovascular magnetic resonance (PC-CMR) can be compromised by background phase errors. It is the objective of the present work to provide an analysis of the temperature dependence of background phase errors in PC-CMR by means of gradient mount temperature sensing and magnetic field monitoring. Methods: Background phase errors were measured for various temperatures of the gradient mount using magnetic field monitoring and validated in a static phantom. The effect of thermal changes during k-space acquisition was simulated and confirmed with measurements in a stationary phantom. Results: The temperature of the gradient mount was found to increase by 20–30 K during PC-CMR measurements of 6–12 min duration. Associated changes in background phase errors of up to 11% or 0.35 radian were measured at 10 cm from the magnet’s iso-center as a result of first order offsets. Zeroth order phase errors exhibited little thermal dependence. Conclusions: It is concluded that changes in gradient mount temperature significantly modify background phase errors during PC-CMR with high gradient duty cycle. Since temperature increases significantly during the first minutes of scanning the results presented are also of relevance for single-slice or multi-slice PC-CMR scans. The findings prompt for further studies to investigate advanced correction methods taking into account gradient temperature and/or the use of concurrent field-monitoring to map gradient-induced fields throughout the scan.

Abstract

Background: The accuracy of phase-contrast cardiovascular magnetic resonance (PC-CMR) can be compromised by background phase errors. It is the objective of the present work to provide an analysis of the temperature dependence of background phase errors in PC-CMR by means of gradient mount temperature sensing and magnetic field monitoring. Methods: Background phase errors were measured for various temperatures of the gradient mount using magnetic field monitoring and validated in a static phantom. The effect of thermal changes during k-space acquisition was simulated and confirmed with measurements in a stationary phantom. Results: The temperature of the gradient mount was found to increase by 20–30 K during PC-CMR measurements of 6–12 min duration. Associated changes in background phase errors of up to 11% or 0.35 radian were measured at 10 cm from the magnet’s iso-center as a result of first order offsets. Zeroth order phase errors exhibited little thermal dependence. Conclusions: It is concluded that changes in gradient mount temperature significantly modify background phase errors during PC-CMR with high gradient duty cycle. Since temperature increases significantly during the first minutes of scanning the results presented are also of relevance for single-slice or multi-slice PC-CMR scans. The findings prompt for further studies to investigate advanced correction methods taking into account gradient temperature and/or the use of concurrent field-monitoring to map gradient-induced fields throughout the scan.

Statistics

Citations

17 citations in Web of Science®
18 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

56 downloads since deposited on 16 Jan 2015
20 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Language:English
Date:2014
Deposited On:16 Jan 2015 09:14
Last Modified:08 Dec 2017 09:51
Publisher:BioMed Central
ISSN:1097-6647
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.1186/s12968-014-0097-6
PubMed ID:25497004

Download

Download PDF  'Analysis of temperature dependence of background phase errors in phase-contrast cardiovascular magnetic resonance'.
Preview
Content: Published Version
Filetype: PDF
Size: 3MB
View at publisher
Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)