Incorporation of time-of-flight information reduces metal artifacts in simultaneous positron emission tomography/magnetic resonance imaging: a simulation study


Davison, Helen; Ter Voert, Edwin E G W; de Galiza Barbosa, Felipe; Veit-Haibach, Patrick; Delso, Gaspar (2015). Incorporation of time-of-flight information reduces metal artifacts in simultaneous positron emission tomography/magnetic resonance imaging: a simulation study. Investigative Radiology, 50(7):423-429.

Abstract

OBJECTIVES: This study aimed to describe and evaluate the influence of time-of-flight (TOF) information on metal artifact reduction in positron emission tomography (PET) image quality in clinical simultaneous PET/magnetic resonance (MR) scanning.
MATERIALS AND METHODS: A total of 7 patients with various malignant tumors were included and underwent a PET/MR examination after standard PET/computed tomography. Baseline TOF and non-TOF PET images were reconstructed. Next, the TOF and non-TOF PET reconstructions were repeated after the introduction of artificial signal voids in the attenuation map to simulate metal artifacts of various sizes in a range of locations. Three different sizes of signal voids were inserted in the attenuation maps for each location of interest: over the maxilla, humeral head, chest, sternum, thoracic and lumbar spine, as well as the femoral head to replicate clinically relevant metal artifacts. The reconstructed images with the artifacts were then compared with the baseline reconstructed images. The mean percentage error in a region of interest surrounding the simulated artifact was used to compare between TOF and non-TOF images. Further comparison between TOF and non-TOF images was performed using histogram analysis.
RESULTS: In all cases, the mean percentage error in a region of interest surrounding the simulated artifact was reduced when TOF information was included in the reconstruction. The inclusion of TOF also changes the distribution of smaller errors away from the origin of the artifact. In some anatomical regions, an increase in the number of small errors was noted with TOF, although the differences with non-TOF were minimal.
CONCLUSIONS: Positron emission tomographic imaging benefits from the integration of TOF information in simultaneous PET/MR. The inclusion of TOF information in simultaneous PET/MR imaging reduces errors related to metal artifacts at the site of the artifact.

Abstract

OBJECTIVES: This study aimed to describe and evaluate the influence of time-of-flight (TOF) information on metal artifact reduction in positron emission tomography (PET) image quality in clinical simultaneous PET/magnetic resonance (MR) scanning.
MATERIALS AND METHODS: A total of 7 patients with various malignant tumors were included and underwent a PET/MR examination after standard PET/computed tomography. Baseline TOF and non-TOF PET images were reconstructed. Next, the TOF and non-TOF PET reconstructions were repeated after the introduction of artificial signal voids in the attenuation map to simulate metal artifacts of various sizes in a range of locations. Three different sizes of signal voids were inserted in the attenuation maps for each location of interest: over the maxilla, humeral head, chest, sternum, thoracic and lumbar spine, as well as the femoral head to replicate clinically relevant metal artifacts. The reconstructed images with the artifacts were then compared with the baseline reconstructed images. The mean percentage error in a region of interest surrounding the simulated artifact was used to compare between TOF and non-TOF images. Further comparison between TOF and non-TOF images was performed using histogram analysis.
RESULTS: In all cases, the mean percentage error in a region of interest surrounding the simulated artifact was reduced when TOF information was included in the reconstruction. The inclusion of TOF also changes the distribution of smaller errors away from the origin of the artifact. In some anatomical regions, an increase in the number of small errors was noted with TOF, although the differences with non-TOF were minimal.
CONCLUSIONS: Positron emission tomographic imaging benefits from the integration of TOF information in simultaneous PET/MR. The inclusion of TOF information in simultaneous PET/MR imaging reduces errors related to metal artifacts at the site of the artifact.

Statistics

Citations

16 citations in Web of Science®
16 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

39 downloads since deposited on 23 Apr 2015
17 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:11 March 2015
Deposited On:23 Apr 2015 10:42
Last Modified:05 Apr 2016 19:13
Publisher:Lippincott Williams & Wilkins
ISSN:0020-9996
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1097/RLI.0000000000000146
PubMed ID:25756682

Download

Preview Icon on Download
Preview
Content: Published Version
Filetype: PDF
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations