Header

UZH-Logo

Maintenance Infos

Diffusion MRI with concurrent magnetic field monitoring


Wilm, Bertram J; Nagy, Zoltan; Barmet, Christoph; Vannesjo, S Johanna; Kasper, Lars; Haeberlin, Max; Gross, Simon; Dietrich, Benjamin E; Brunner, David O; Schmid, Thomas; Pruessmann, Klaas P (2015). Diffusion MRI with concurrent magnetic field monitoring. Magnetic Resonance in Medicine, 74(4):925-933.

Abstract

PURPOSE: Diffusion MRI is compromised by unknown field perturbation during image encoding. The purpose of this study was to address this problem using the recently described approach of concurrent magnetic field monitoring.
METHODS: Magnetic field dynamics were monitored during the echo planar imaging readout of a common diffusion-weighted MRI sequence using an integrated magnetic field camera setup. The image encoding including encoding changes over the duration of entire scans were quantified and analyzed. Field perturbations were corrected by accounting for them in generalized image reconstruction. The impact on image quality along with geometrical congruence among different diffusion-weighted images was assessed both qualitatively and quantitatively.
RESULTS: The most significant field perturbations were found to be related to higher-order eddy currents from diffusion-weighting gradients and B0 field drift as well as gradual changes of short-term eddy current behavior and mechanical oscillations during the scan. All artifacts relating to dynamic field perturbations were eliminated by incorporating the measured encoding in image reconstruction.
CONCLUSION: Concurrent field monitoring combined with generalized reconstruction enhances depiction fidelity in diffusion imaging. In addition to artifact reduction, it improves geometric congruence and thus facilitates image combination for quantitative diffusion analysis.

Abstract

PURPOSE: Diffusion MRI is compromised by unknown field perturbation during image encoding. The purpose of this study was to address this problem using the recently described approach of concurrent magnetic field monitoring.
METHODS: Magnetic field dynamics were monitored during the echo planar imaging readout of a common diffusion-weighted MRI sequence using an integrated magnetic field camera setup. The image encoding including encoding changes over the duration of entire scans were quantified and analyzed. Field perturbations were corrected by accounting for them in generalized image reconstruction. The impact on image quality along with geometrical congruence among different diffusion-weighted images was assessed both qualitatively and quantitatively.
RESULTS: The most significant field perturbations were found to be related to higher-order eddy currents from diffusion-weighting gradients and B0 field drift as well as gradual changes of short-term eddy current behavior and mechanical oscillations during the scan. All artifacts relating to dynamic field perturbations were eliminated by incorporating the measured encoding in image reconstruction.
CONCLUSION: Concurrent field monitoring combined with generalized reconstruction enhances depiction fidelity in diffusion imaging. In addition to artifact reduction, it improves geometric congruence and thus facilitates image combination for quantitative diffusion analysis.

Statistics

Citations

Dimensions.ai Metrics
8 citations in Web of Science®
8 citations in Scopus®
8 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

33 downloads since deposited on 17 Nov 2015
18 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Uncontrolled Keywords:DTI; DWI; artifact correction; diffusion; distortion correction; field monitoring
Language:English
Date:2015
Deposited On:17 Nov 2015 10:49
Last Modified:14 Feb 2018 09:43
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:0740-3194
OA Status:Green
Publisher DOI:https://doi.org/10.1002/mrm.25827
PubMed ID:26183218