Comparison of conventional ROTEM® cups and pins to the ROTEM® cup and pin mini measuring cells (MiniCup)


Haas, Thorsten; Spielmann, Nelly; Dillier, Claudia; Cushing, Melissa; Siegmund, Svenja; Krüger, Bernard (2015). Comparison of conventional ROTEM® cups and pins to the ROTEM® cup and pin mini measuring cells (MiniCup). Scandinavian Journal of Clinical and Laboratory Investigation, 75(6):470-475.

Abstract

BACKGROUND: Thromboelastometry (ROTEM®) is a well-established measurement to guide perioperative coagulation management. Recently, an enhancement for low volume specimens using cup and pin mini measuring cells (MiniCup) was invented for scientific non-clinical use. The MiniCup measuring cells allow testing with half of the blood volume required to run a test as compared to the conventional ROTEM® cup and pin (150 μL vs. 300 μL per assay).
OBJECTIVES: The aim of this prospective study was to compare ROTEM® results using conventional cups and pins to the results of the MiniCup measuring cells.
METHODS: Blood samples drawn during major pediatric surgery were analyzed using the conventional ROTEM® and the MiniCup system to compare parameters from the EXTEM, INTEM and FIBTEM assays.
RESULTS: Citrated blood samples (n = 120) from 70 pediatric patients were analyzed. Results of the MiniCup cells were considerably different than the conventional ROTEM® cups and pins measurements. The MiniCup results show less clot firmness and demonstrate higher variability in the parameters reflecting the kinetics of clot building.
CONCLUSION: The MiniCup measuring cells may offer advantages in pediatric care or research facilities, but specific reference ranges need to be established first, and adequate reproducibility must be determined by further studies before clinical use can be recommended.

Abstract

BACKGROUND: Thromboelastometry (ROTEM®) is a well-established measurement to guide perioperative coagulation management. Recently, an enhancement for low volume specimens using cup and pin mini measuring cells (MiniCup) was invented for scientific non-clinical use. The MiniCup measuring cells allow testing with half of the blood volume required to run a test as compared to the conventional ROTEM® cup and pin (150 μL vs. 300 μL per assay).
OBJECTIVES: The aim of this prospective study was to compare ROTEM® results using conventional cups and pins to the results of the MiniCup measuring cells.
METHODS: Blood samples drawn during major pediatric surgery were analyzed using the conventional ROTEM® and the MiniCup system to compare parameters from the EXTEM, INTEM and FIBTEM assays.
RESULTS: Citrated blood samples (n = 120) from 70 pediatric patients were analyzed. Results of the MiniCup cells were considerably different than the conventional ROTEM® cups and pins measurements. The MiniCup results show less clot firmness and demonstrate higher variability in the parameters reflecting the kinetics of clot building.
CONCLUSION: The MiniCup measuring cells may offer advantages in pediatric care or research facilities, but specific reference ranges need to be established first, and adequate reproducibility must be determined by further studies before clinical use can be recommended.

Statistics

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery
Dewey Decimal Classification:610 Medicine & health
Uncontrolled Keywords:Blood coagulation; pediatric anesthesia; point-of-care testing; thrombelastography; thromboelastometry
Language:English
Date:2015
Deposited On:30 Nov 2015 10:58
Last Modified:08 Dec 2017 15:14
Publisher:Taylor & Francis
ISSN:0036-5513
Publisher DOI:https://doi.org/10.3109/00365513.2015.1042909
PubMed ID:25990261

Download

Full text not available from this repository.
View at publisher