A g-factor metric for k-t SENSE and k-t PCA based parallel imaging


Binter, Christian; Ramb, Rebecca; Jung, Bernd; Kozerke, Sebastian (2016). A g-factor metric for k-t SENSE and k-t PCA based parallel imaging. Magnetic Resonance in Medicine, 75(2):562-571.

Abstract

PURPOSE: To propose and validate a g-factor formalism for k-t SENSE, k-t PCA and related k-t methods for assessing SNR and temporal fidelity.
METHODS: An analytical gxf -factor formulation in the spatiotemporal frequency domain is derived, enabling assessment of noise and depiction fidelity in both the spatial and frequency domain. Using pseudoreplica analysis of cardiac cine data the gxf -factor description is validated and example data are used to analyze the performance of k-t methods for various parameter settings.
RESULTS: Analytical gxf -factor maps were found to agree well with pseudoreplica analysis for 3x, 5x, and 7x k-t SENSE and k-t PCA. While k-t SENSE resulted in lower average gxf values (gx (avg) ) in static regions when compared with k-t PCA, k-t PCA yielded lower gx (avg) values in dynamic regions. Temporal transfer was better preserved with k-t PCA for increasing undersampling factors.
CONCLUSION: The proposed gxf -factor and temporal transfer formalism allows assessing noise performance and temporal depiction fidelity of k-t methods including k-t SENSE and k-t PCA. The framework enables quantitative comparison of different k-t methods relative to frame-by-frame parallel imaging reconstruction. Magn Reson Med 75:562-571, 2016. © 2015 Wiley Periodicals, Inc.

Abstract

PURPOSE: To propose and validate a g-factor formalism for k-t SENSE, k-t PCA and related k-t methods for assessing SNR and temporal fidelity.
METHODS: An analytical gxf -factor formulation in the spatiotemporal frequency domain is derived, enabling assessment of noise and depiction fidelity in both the spatial and frequency domain. Using pseudoreplica analysis of cardiac cine data the gxf -factor description is validated and example data are used to analyze the performance of k-t methods for various parameter settings.
RESULTS: Analytical gxf -factor maps were found to agree well with pseudoreplica analysis for 3x, 5x, and 7x k-t SENSE and k-t PCA. While k-t SENSE resulted in lower average gxf values (gx (avg) ) in static regions when compared with k-t PCA, k-t PCA yielded lower gx (avg) values in dynamic regions. Temporal transfer was better preserved with k-t PCA for increasing undersampling factors.
CONCLUSION: The proposed gxf -factor and temporal transfer formalism allows assessing noise performance and temporal depiction fidelity of k-t methods including k-t SENSE and k-t PCA. The framework enables quantitative comparison of different k-t methods relative to frame-by-frame parallel imaging reconstruction. Magn Reson Med 75:562-571, 2016. © 2015 Wiley Periodicals, Inc.

Statistics

Citations

3 citations in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 03 Feb 2016
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Uncontrolled Keywords:g-factor; k-t PCA; k-t SENSE; noise estimation; parallel imaging; temporal filtering
Language:English
Date:2016
Deposited On:03 Feb 2016 15:07
Last Modified:23 Nov 2017 14:29
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:0740-3194
Publisher DOI:https://doi.org/10.1002/mrm.25606
PubMed ID:25809284

Download