Header

UZH-Logo

Maintenance Infos

Safety testing and operational procedures for self-developed radiofrequency coils


Hoffmann, Jens; Henning, Anke; Giapitzakis, Ioannis A; Scheffler, Klaus; Shajan, G; Pohmann, Rolf; Avdievich, Nikolai I (2016). Safety testing and operational procedures for self-developed radiofrequency coils. NMR in Biomedicine, 29(9):1131-1144.

Abstract

The development of novel radiofrequency (RF) coils for human ultrahigh-field (≥7 T), non-proton and body applications is an active field of research in many MR groups. Any RF coil must meet the strict requirements for safe application on humans with respect to mechanical and electrical safety, as well as the specific absorption rate (SAR) limits. For this purpose, regulations such as the International Electrotechnical Commission (IEC) standard for medical electrical equipment, vendor-suggested test specifications for third party coils and custom-developed test procedures exist. However, for higher frequencies and shorter wavelengths in ultrahigh-field MR, the RF fields may become extremely inhomogeneous in biological tissue and the risk of localized areas with elevated power deposition increases, which is usually not considered by existing safety testing and operational procedures. In addition, important aspects, such as risk analysis and comprehensive electrical performance and safety tests, are often neglected. In this article, we describe the guidelines used in our institution for electrical and mechanical safety tests, SAR simulation and verification, risk analysis and operational procedures, including coil documentation, user training and regular quality assurance testing, which help to recognize and eliminate safety issues during coil design and operation. Although the procedure is generally applicable to all field strengths, specific requirements with regard to SAR-related safety and electrical performance at ultrahigh-field are considered. The protocol describes an internal procedure and does not reflect consensus among a large number of research groups, but rather aims to stimulate further discussion related to minimum coil safety standards. Furthermore, it may help other research groups to establish their own procedures. Copyright © 2015 John Wiley & Sons, Ltd.

Abstract

The development of novel radiofrequency (RF) coils for human ultrahigh-field (≥7 T), non-proton and body applications is an active field of research in many MR groups. Any RF coil must meet the strict requirements for safe application on humans with respect to mechanical and electrical safety, as well as the specific absorption rate (SAR) limits. For this purpose, regulations such as the International Electrotechnical Commission (IEC) standard for medical electrical equipment, vendor-suggested test specifications for third party coils and custom-developed test procedures exist. However, for higher frequencies and shorter wavelengths in ultrahigh-field MR, the RF fields may become extremely inhomogeneous in biological tissue and the risk of localized areas with elevated power deposition increases, which is usually not considered by existing safety testing and operational procedures. In addition, important aspects, such as risk analysis and comprehensive electrical performance and safety tests, are often neglected. In this article, we describe the guidelines used in our institution for electrical and mechanical safety tests, SAR simulation and verification, risk analysis and operational procedures, including coil documentation, user training and regular quality assurance testing, which help to recognize and eliminate safety issues during coil design and operation. Although the procedure is generally applicable to all field strengths, specific requirements with regard to SAR-related safety and electrical performance at ultrahigh-field are considered. The protocol describes an internal procedure and does not reflect consensus among a large number of research groups, but rather aims to stimulate further discussion related to minimum coil safety standards. Furthermore, it may help other research groups to establish their own procedures. Copyright © 2015 John Wiley & Sons, Ltd.

Statistics

Citations

10 citations in Web of Science®
6 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 03 Feb 2016
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Uncontrolled Keywords:Numerical Simulations; RF Safety; RF receive coils; RF transmit coils; SAR; Ultrahigh-field
Language:English
Date:2016
Deposited On:03 Feb 2016 15:07
Last Modified:08 Dec 2017 17:39
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:0952-3480
Publisher DOI:https://doi.org/10.1002/nbm.3290
PubMed ID:25851551

Download