Multicentric performance analysis of HCV quantification assays and its potential relevance for HCV treatment


Wiesmann, F; Naeth, G; Berger, A; Hirsch, H H; Regenass, S; Ross, R S; Sarrazin, C; Wedemeyer, H; Knechten, H; Braun, P (2016). Multicentric performance analysis of HCV quantification assays and its potential relevance for HCV treatment. Medical Microbiology and Immunology, 205(3):263-268.

Abstract

An accurate quantification of low viremic HCV RNA plasma samples has gained importance since the approval of direct acting antivirals and since only one single measurement predicts the necessity of a prolonged or shortened therapy. As reported previously, HCV quantification assays such as Abbott RealTime HCV and Roche COBAS AmpliPrep/COBAS TaqMan HCV version 2 (CTM v2) may vary in sensitivity and precision particularly in low-level viremia. Importantly, substantial variations were previously demonstrated between some of these assays compared to the Roche High Pure System/COBAS TaqMan assay (HPS) reference assay, which was used to establish the clinical decision points in clinical studies. In this study, the reproducibility of assay performances across several laboratories was assessed by analysing quantification results generated by six independent laboratories (3× RealTime, 3× CTM v2) in comparison with one HPS reference laboratory. The 4th WHO Standard was diluted to 100, 25 and 10 IU/ml, and aliquots were tested in triplicates in 5 independent runs by each assay in the different laboratories to assess assay precision and detection rates. In a second approach, 2 clinical samples (GT 1a & GT 1b) were diluted to 100 and 25 IU/ml and tested as described above. While the result range for WHO 100 IU/ml replicates across all laboratories was similar in this analysis, the CVs of each laboratory ranged from 19.3 to 25.6 % for RealTime laboratories and were lower than CVs of CTM v2 laboratories with a range of 26.1-47.3 %, respectively, and also in comparison with the CV of the HPS reference laboratory (34.9 %). At WHO standard dilution of 25 IU/ml, 24 replicates were quantified by RealTime compared to 8 replicates with CTM v2. Results of clinical samples again revealed a higher variation of CTM v2 results as compared to RealTime values. (CVs at 100 IU/ml: RealTime: 13.1-21.0 % and CTM v2: 15.0-32.3 %; CVs at 25 IU/ml: RealTime 17.6-34.9 % and CTM v2 28.2-54.9 %). These findings confirm the superior precision of RealTime versus CTM v2 at low-level viremia even across different laboratories including the new clinical decision point at 25 IU/ml. A highly precise monitoring of HCV viral load during therapy will remain crucial for patient management with regard to futility rules, therapy efficacy and SVR.

Abstract

An accurate quantification of low viremic HCV RNA plasma samples has gained importance since the approval of direct acting antivirals and since only one single measurement predicts the necessity of a prolonged or shortened therapy. As reported previously, HCV quantification assays such as Abbott RealTime HCV and Roche COBAS AmpliPrep/COBAS TaqMan HCV version 2 (CTM v2) may vary in sensitivity and precision particularly in low-level viremia. Importantly, substantial variations were previously demonstrated between some of these assays compared to the Roche High Pure System/COBAS TaqMan assay (HPS) reference assay, which was used to establish the clinical decision points in clinical studies. In this study, the reproducibility of assay performances across several laboratories was assessed by analysing quantification results generated by six independent laboratories (3× RealTime, 3× CTM v2) in comparison with one HPS reference laboratory. The 4th WHO Standard was diluted to 100, 25 and 10 IU/ml, and aliquots were tested in triplicates in 5 independent runs by each assay in the different laboratories to assess assay precision and detection rates. In a second approach, 2 clinical samples (GT 1a & GT 1b) were diluted to 100 and 25 IU/ml and tested as described above. While the result range for WHO 100 IU/ml replicates across all laboratories was similar in this analysis, the CVs of each laboratory ranged from 19.3 to 25.6 % for RealTime laboratories and were lower than CVs of CTM v2 laboratories with a range of 26.1-47.3 %, respectively, and also in comparison with the CV of the HPS reference laboratory (34.9 %). At WHO standard dilution of 25 IU/ml, 24 replicates were quantified by RealTime compared to 8 replicates with CTM v2. Results of clinical samples again revealed a higher variation of CTM v2 results as compared to RealTime values. (CVs at 100 IU/ml: RealTime: 13.1-21.0 % and CTM v2: 15.0-32.3 %; CVs at 25 IU/ml: RealTime 17.6-34.9 % and CTM v2 28.2-54.9 %). These findings confirm the superior precision of RealTime versus CTM v2 at low-level viremia even across different laboratories including the new clinical decision point at 25 IU/ml. A highly precise monitoring of HCV viral load during therapy will remain crucial for patient management with regard to futility rules, therapy efficacy and SVR.

Statistics

Citations

6 citations in Web of Science®
6 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 29 Feb 2016
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Immunology
Dewey Decimal Classification:610 Medicine & health
Uncontrolled Keywords:Assay variation; COBAS Taqman; HCV; RealTime; Viral load
Language:English
Date:2016
Deposited On:29 Feb 2016 09:55
Last Modified:14 May 2016 01:03
Publisher:Springer
ISSN:0300-8584
Publisher DOI:https://doi.org/10.1007/s00430-015-0443-9
PubMed ID:26666643

Download

Preview Icon on Download
Filetype: PDF - Registered users only
Size: 850kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations