Quantitative comparison of 2D and 3D late gadolinium enhancement MR imaging in patients with Fabry disease and hypertrophic cardiomyopathy


Morsbach, F; Gordic, S; Gruner, C; Niemann, M; Goetti, R; Gotschy, A; Kozerke, S; Alkadhi, H; Manka, Robert (2016). Quantitative comparison of 2D and 3D late gadolinium enhancement MR imaging in patients with Fabry disease and hypertrophic cardiomyopathy. International Journal of Cardiology, 217:167-173.

Abstract

BACKGROUND: This study aims to determine whether the quantification of myocardial fibrosis in patients with Fabry disease (FD) and hypertrophic cardiomyopathy (HCM) using a late gadolinium enhancement (LGE) singlebreath-hold three-dimensional (3D) inversion recovery magnetic resonance (MR) imaging sequence is comparable with a clinically established two-dimensional (2D) multi-breath-hold sequence.
METHODS: In this retrospective, IRB-approved study, 40 consecutive patients (18 male; mean age 50±17years) with Fabry disease (n=18) and HCM (n=22) underwent MR imaging at 1.5T. Spatial resolution was the same for 3D and 2D images (field-of-view, 350×350mm(2); in-plane-resolution, 1.2×1.2mm(2); section-thickness, 8mm). Datasets were analyzed for subjective image quality; myocardial and fibrotic mass, and total fibrotic tissue percentage were quantified.
RESULTS: There was no significant difference in subjective image quality between 3D and 2D acquisitions (P=0.1 and P=0.3) for either disease. In patients with Fabry disease there were no significant differences between 3D and 2D acquisitions for myocardial mass (P=0.55), fibrous tissue mass (P=0.89), and total fibrous percentage (P=0.67), with good agreement between acquisitions according to Bland-Altman analyses. In patients with HCM there were also no significant differences between acquisitions for myocardial mass (P=0.48), fibrous tissue mass (P=0.56), and total fibrous percentage (P=0.67), with good agreement according to Bland-Altman analyses. Acquisition time was significantly shorter for 3D (25±5s) as compared to the 2D sequence (349±62s, P<0.001).
CONCLUSIONS: In patients with Fabry disease and HCM, 3D LGE imaging provides equivalent diagnostic information in regard to quantification of myocardial fibrosis as compared with a standard 2D sequence, but at superior acquisition speed.

Abstract

BACKGROUND: This study aims to determine whether the quantification of myocardial fibrosis in patients with Fabry disease (FD) and hypertrophic cardiomyopathy (HCM) using a late gadolinium enhancement (LGE) singlebreath-hold three-dimensional (3D) inversion recovery magnetic resonance (MR) imaging sequence is comparable with a clinically established two-dimensional (2D) multi-breath-hold sequence.
METHODS: In this retrospective, IRB-approved study, 40 consecutive patients (18 male; mean age 50±17years) with Fabry disease (n=18) and HCM (n=22) underwent MR imaging at 1.5T. Spatial resolution was the same for 3D and 2D images (field-of-view, 350×350mm(2); in-plane-resolution, 1.2×1.2mm(2); section-thickness, 8mm). Datasets were analyzed for subjective image quality; myocardial and fibrotic mass, and total fibrotic tissue percentage were quantified.
RESULTS: There was no significant difference in subjective image quality between 3D and 2D acquisitions (P=0.1 and P=0.3) for either disease. In patients with Fabry disease there were no significant differences between 3D and 2D acquisitions for myocardial mass (P=0.55), fibrous tissue mass (P=0.89), and total fibrous percentage (P=0.67), with good agreement between acquisitions according to Bland-Altman analyses. In patients with HCM there were also no significant differences between acquisitions for myocardial mass (P=0.48), fibrous tissue mass (P=0.56), and total fibrous percentage (P=0.67), with good agreement according to Bland-Altman analyses. Acquisition time was significantly shorter for 3D (25±5s) as compared to the 2D sequence (349±62s, P<0.001).
CONCLUSIONS: In patients with Fabry disease and HCM, 3D LGE imaging provides equivalent diagnostic information in regard to quantification of myocardial fibrosis as compared with a standard 2D sequence, but at superior acquisition speed.

Statistics

Citations

Dimensions.ai Metrics
3 citations in Web of Science®
3 citations in Scopus®
3 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 26 May 2016
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Date:3 May 2016
Deposited On:26 May 2016 19:48
Last Modified:02 Feb 2018 10:01
Publisher:Elsevier
ISSN:0167-5273
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.ijcard.2016.04.175
PubMed ID:27183453

Download