Registration accuracy of CT/MRI fusion for localisation of deep brain stimulation electrode position: an imaging study and systematic review


Geevarghese, Ruben; O'Gorman Tuura, Ruth; Lumsden, Daniel E; Samuel, Michael; Ashkan, Keyoumars (2016). Registration accuracy of CT/MRI fusion for localisation of deep brain stimulation electrode position: an imaging study and systematic review. Stereotactic and functional neurosurgery, 94(3):159-163.

Abstract

BACKGROUND: Postoperative imaging is essential for verifying electrode location in patients undergoing deep brain stimulation (DBS). MRI offers better visualisation of brain targets, but concerns about adverse events have limited its use. Preoperative stereotactic MRI fused with a postoperative stereotactic CT, demonstrating the electrode position, is now widely used.
OBJECTIVES: The aims of this study were to: (1) evaluate the accuracy of image registration using Neuroinspire, and (2) undertake a systematic review of the literature on CT/MRI fusion techniques to ascertain the accuracy of other software packages.
METHODS: Twenty patients who underwent bilateral subthalamic nucleus DBS for Parkinson's disease were selected. The postoperative CT was registered and fused with the preoperative MRI using Neuroinspire. The position of each electrode tip was determined in stereotactic coordinates both in the (unfused) postoperative CT and the fused CT/MRI. The difference in tip position was used to evaluate the registration accuracy.
RESULTS: The mean error ± SD of CT/MRI fusion using Neuroinspire was 0.25 ± 0.15, 0.33 ± 0.26 and 0.46 ± 0.55 mm in lateral, anteroposterior and vertical axes. A systematic review suggested that CT/MRI registration with Neuroinspire is more accurate than that achieved with other tested CT/MRI fusion algorithms.
CONCLUSION: CT/MRI fusion for localisation of electrode placement offers an accurate, reliable and safe modality for assessing electrode location.

Abstract

BACKGROUND: Postoperative imaging is essential for verifying electrode location in patients undergoing deep brain stimulation (DBS). MRI offers better visualisation of brain targets, but concerns about adverse events have limited its use. Preoperative stereotactic MRI fused with a postoperative stereotactic CT, demonstrating the electrode position, is now widely used.
OBJECTIVES: The aims of this study were to: (1) evaluate the accuracy of image registration using Neuroinspire, and (2) undertake a systematic review of the literature on CT/MRI fusion techniques to ascertain the accuracy of other software packages.
METHODS: Twenty patients who underwent bilateral subthalamic nucleus DBS for Parkinson's disease were selected. The postoperative CT was registered and fused with the preoperative MRI using Neuroinspire. The position of each electrode tip was determined in stereotactic coordinates both in the (unfused) postoperative CT and the fused CT/MRI. The difference in tip position was used to evaluate the registration accuracy.
RESULTS: The mean error ± SD of CT/MRI fusion using Neuroinspire was 0.25 ± 0.15, 0.33 ± 0.26 and 0.46 ± 0.55 mm in lateral, anteroposterior and vertical axes. A systematic review suggested that CT/MRI registration with Neuroinspire is more accurate than that achieved with other tested CT/MRI fusion algorithms.
CONCLUSION: CT/MRI fusion for localisation of electrode placement offers an accurate, reliable and safe modality for assessing electrode location.

Statistics

Citations

Dimensions.ai Metrics
3 citations in Web of Science®
4 citations in Scopus®
4 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

29 downloads since deposited on 23 Dec 2016
29 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Medical Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:18 June 2016
Deposited On:23 Dec 2016 10:28
Last Modified:02 Feb 2018 10:06
Publisher:Karger
ISSN:1011-6125
OA Status:Green
Publisher DOI:https://doi.org/10.1159/000446609
PubMed ID:27318464

Download

Download PDF  'Registration accuracy of CT/MRI fusion for localisation of deep brain stimulation electrode position: an imaging study and systematic review'.
Preview
Content: Published Version
Filetype: PDF
Size: 130kB
View at publisher