Validation of the Accutrend Plus point-of-care triglyceride analyzer in horses, ponies, and donkeys


Pongratz, Michaela Carina; Junge, Hannah K; Riond, Barbara; Schwarzwald, Colin C (2016). Validation of the Accutrend Plus point-of-care triglyceride analyzer in horses, ponies, and donkeys. Journal of Veterinary Emergency and Critical Care, 26(5):682-690.

Abstract

OBJECTIVE: To assess the accuracy and reliability of a point-of-care (POC) triglyceride analyzer and to establish reference intervals for blood ([TRIG]POC/WB ) and plasma triglyceride concentrations ([TRIG]POC/PL ) in horses, ponies, and donkeys.
DESIGN: Prospective study.
SETTING: University teaching hospital.
ANIMALS: 120 adult healthy equids (78 horses and ponies, 42 donkeys) and 79 equids suffering from hypertriglyceridemia (73 horses and ponies, 6 donkeys).
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: [TRIG]POC/WB and [TRIG]POC/PL were measured using a POC analyzer and plasma triglyceride concentrations were measured using a standard laboratory assay ([TRIG]LAB/PL ). Reference intervals were determined. Test accuracy was assessed by Bland-Altman comparison of POC measurements with standard laboratory measurements and by evaluating linearity of dilutional series. Test reliability was assessed by repeated serial measurements. [TRIG]POC/WB and [TRIG]POC/PL were below the analytic range of the POC assay (<0.8 mmol/L [<70 mg/dL]) in healthy horses and ponies, whereas the reference intervals were 0.82-3.14 mmol/L (73-278 mg/dL) and 0.87-3.02 mmol/L (77-267 mg/dL), respectively, in donkeys. The POC analyzer systematically overestimated triglyceride concentrations when compared to a standard laboratory assay. The difference between [TRIG]POC/WB and [TRIG]POC/PL was small and clinically negligible. The coefficient of variation of repeated measures performed on the POC analyzer was below 10% for [TRIG]POC/WB and [TRIG]POC/PL , both in horses and donkeys and at all concentration ranges.
CONCLUSIONS: The POC analyzer allows accurate and reliable measurement of [TRIG]POC/WB and [TRIG]POC/PL in horses, ponies, and donkeys in clinical settings. Assay-specific reference intervals should be determined for diagnosis and clinical monitoring of hypertriglyceridemia in equids.

Abstract

OBJECTIVE: To assess the accuracy and reliability of a point-of-care (POC) triglyceride analyzer and to establish reference intervals for blood ([TRIG]POC/WB ) and plasma triglyceride concentrations ([TRIG]POC/PL ) in horses, ponies, and donkeys.
DESIGN: Prospective study.
SETTING: University teaching hospital.
ANIMALS: 120 adult healthy equids (78 horses and ponies, 42 donkeys) and 79 equids suffering from hypertriglyceridemia (73 horses and ponies, 6 donkeys).
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: [TRIG]POC/WB and [TRIG]POC/PL were measured using a POC analyzer and plasma triglyceride concentrations were measured using a standard laboratory assay ([TRIG]LAB/PL ). Reference intervals were determined. Test accuracy was assessed by Bland-Altman comparison of POC measurements with standard laboratory measurements and by evaluating linearity of dilutional series. Test reliability was assessed by repeated serial measurements. [TRIG]POC/WB and [TRIG]POC/PL were below the analytic range of the POC assay (<0.8 mmol/L [<70 mg/dL]) in healthy horses and ponies, whereas the reference intervals were 0.82-3.14 mmol/L (73-278 mg/dL) and 0.87-3.02 mmol/L (77-267 mg/dL), respectively, in donkeys. The POC analyzer systematically overestimated triglyceride concentrations when compared to a standard laboratory assay. The difference between [TRIG]POC/WB and [TRIG]POC/PL was small and clinically negligible. The coefficient of variation of repeated measures performed on the POC analyzer was below 10% for [TRIG]POC/WB and [TRIG]POC/PL , both in horses and donkeys and at all concentration ranges.
CONCLUSIONS: The POC analyzer allows accurate and reliable measurement of [TRIG]POC/WB and [TRIG]POC/PL in horses, ponies, and donkeys in clinical settings. Assay-specific reference intervals should be determined for diagnosis and clinical monitoring of hypertriglyceridemia in equids.

Statistics

Altmetrics

Downloads

0 downloads since deposited on 24 Aug 2016
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Equine Department
05 Vetsuisse Faculty > Veterinary Clinic > Department of Farm Animals
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Date:2016
Deposited On:24 Aug 2016 14:31
Last Modified:13 Jan 2017 09:02
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:1476-4431
Publisher DOI:https://doi.org/10.1111/vec.12512
PubMed ID:27508527

Download

Preview Icon on Download
Content: Published Version
Language: English
Filetype: PDF - Registered users only
Size: 315kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations