Determination of reference intervals and comparison of venous blood gas parameters using standard and non-standard collection methods in 24 cats


Bachmann, Karin; Kutter, Annette P N; Schefer, R J; Marly-Voquer, C; Sigrist, Nadja (2017). Determination of reference intervals and comparison of venous blood gas parameters using standard and non-standard collection methods in 24 cats. Journal of Feline Medicine and Surgery, 19(8):831-840.

Abstract

OBJECTIVES: The aim of this study was to determine in-house reference intervals (RIs) for venous blood analysis with the RAPIDPoint 500 blood gas analyser using blood gas syringes (BGSs) and to determine whether immediate analysis of venous blood collected into lithium heparin (LH) tubes can replace anaerobic blood sampling into BGSs.
METHODS: Venous blood was collected from 24 healthy cats and directly transferred into a BGS and a LH tube. The BGS was immediately analysed on the RAPIDPoint 500 followed by the LH tube. The BGSs and LH tubes were compared using paired t-test or Wilcoxon matched-pairs signed-rank test, Bland-Altman and Passing-Bablok analysis. To assess clinical relevance, bias or percentage bias between BGSs and LH tubes was compared with the allowable total error (TEa) recommended for the respective parameter.
RESULTS: Based on the values obtained from the BGSs, RIs were calculated for the evaluated parameters, including blood gases, electrolytes, glucose and lactate. Values derived from LH tubes showed no significant difference for standard bicarbonate, whole blood base excess, haematocrit, total haemoglobin, sodium, potassium, chloride, glucose and lactate, while pH, partial pressure of carbon dioxide and oxygen, actual bicarbonate, extracellular base excess, ionised calcium and anion gap were significantly different to the samples collected in BGSs (P <0.05). Furthermore, pH, partial pressure of carbon dioxide and oxygen, extracellular base excess, ionised calcium and anion gap exceeded the recommended TEa.
CONCLUSIONS AND RELEVANCE: Assessment of actual and standard bicarbonate, whole blood base excess, haematocrit, total haemoglobin, sodium, potassium, chloride, glucose and lactate can be made based on blood collected in LH tubes and analysed within 5 mins. For pH, partial pressure of carbon dioxide and oxygen, extracellular base excess, anion gap and ionised calcium the clinically relevant alterations have to be considered if analysed in LH tubes

Abstract

OBJECTIVES: The aim of this study was to determine in-house reference intervals (RIs) for venous blood analysis with the RAPIDPoint 500 blood gas analyser using blood gas syringes (BGSs) and to determine whether immediate analysis of venous blood collected into lithium heparin (LH) tubes can replace anaerobic blood sampling into BGSs.
METHODS: Venous blood was collected from 24 healthy cats and directly transferred into a BGS and a LH tube. The BGS was immediately analysed on the RAPIDPoint 500 followed by the LH tube. The BGSs and LH tubes were compared using paired t-test or Wilcoxon matched-pairs signed-rank test, Bland-Altman and Passing-Bablok analysis. To assess clinical relevance, bias or percentage bias between BGSs and LH tubes was compared with the allowable total error (TEa) recommended for the respective parameter.
RESULTS: Based on the values obtained from the BGSs, RIs were calculated for the evaluated parameters, including blood gases, electrolytes, glucose and lactate. Values derived from LH tubes showed no significant difference for standard bicarbonate, whole blood base excess, haematocrit, total haemoglobin, sodium, potassium, chloride, glucose and lactate, while pH, partial pressure of carbon dioxide and oxygen, actual bicarbonate, extracellular base excess, ionised calcium and anion gap were significantly different to the samples collected in BGSs (P <0.05). Furthermore, pH, partial pressure of carbon dioxide and oxygen, extracellular base excess, ionised calcium and anion gap exceeded the recommended TEa.
CONCLUSIONS AND RELEVANCE: Assessment of actual and standard bicarbonate, whole blood base excess, haematocrit, total haemoglobin, sodium, potassium, chloride, glucose and lactate can be made based on blood collected in LH tubes and analysed within 5 mins. For pH, partial pressure of carbon dioxide and oxygen, extracellular base excess, anion gap and ionised calcium the clinically relevant alterations have to be considered if analysed in LH tubes

Statistics

Altmetrics

Downloads

22 downloads since deposited on 25 Aug 2016
22 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Equine Department
05 Vetsuisse Faculty > Veterinary Clinic > Department of Small Animals
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Date:2017
Deposited On:25 Aug 2016 12:41
Last Modified:26 Jul 2017 01:00
Publisher:Sage Publications Ltd.
ISSN:1098-612X
Publisher DOI:https://doi.org/10.1177/1098612X16663269
PubMed ID:27542821

Download

Preview Icon on Download
Content: Published Version
Filetype: PDF - Registered users only
Size: 2MB
View at publisher
Preview Icon on Download
Preview
Content: Accepted Version
Language: English
Filetype: PDF
Size: 407kB