Header

UZH-Logo

Maintenance Infos

Dose Reduction in Tomosynthesis of the Wrist


Becker, Anton S; Martini, Katharina; Higashigaito, Kai; Guggenberger, Roman; Andreisek, Gustav; Frauenfelder, Thomas (2016). Dose Reduction in Tomosynthesis of the Wrist. American Journal of Roentgenology:1-6.

Abstract

OBJECTIVE The purpose of this study was to quantitatively and qualitatively determine the impact of radiation dose reduction on the image noise and quality of tomosynthesis studies of the wrist. MATERIALS AND METHODS Imaging of six cadaver wrists was performed with tomosynthesis in anteroposterior position at a tube voltage of 60 kV and tube current of 80 mA and subsequently at 60 or 50 kV with different tube currents of 80, 40, or 32 mA. Dose-area products (DAP) were obtained from the electronically logged protocol. Image noise was measured with an ROI. Two independent and blinded readers evaluated all images. Interreader agreement was measured with a Cohen kappa. Readers assessed overall quality and delineation of soft tissue, cortical bone, and trabecular bone on a 4-point Likert scale. RESULTS The highest DAP (3.892 ± 0.432 Gy · cm(2)) was recorded for images obtained with 60 kV and 80 mA; the lowest (0.857 ± 0.178 Gy · cm(2)) was recorded for images obtained with 50 kV and 32 mA. Noise was highest when a combination of 50 kV and 32 mA (389 ± 26.6) was used and lowest when a combination of 60 kV and 80 mA (218 ± 12.3) was used. The amount of noise on images acquired using 60 kV and 80 mA was statistically significantly different from the amount measured on all other images (p < 0.0001). Interreader agreement was excellent (κ = 0.93). Delineation of anatomy and overall quality were scored best on images obtained with 60 kV and 80 mA and worst on images obtained with 50 kV and 32 mA. The difference in delineation and quality on images obtained using 50 kV and 40 mA was not statistically significantly different compared with images obtained using 60 kV and 80 mA. CONCLUSION Significant dose reduction for tomosynthesis of the wrist is possible while image quality and delineation of anatomic structures remain preserved.

Abstract

OBJECTIVE The purpose of this study was to quantitatively and qualitatively determine the impact of radiation dose reduction on the image noise and quality of tomosynthesis studies of the wrist. MATERIALS AND METHODS Imaging of six cadaver wrists was performed with tomosynthesis in anteroposterior position at a tube voltage of 60 kV and tube current of 80 mA and subsequently at 60 or 50 kV with different tube currents of 80, 40, or 32 mA. Dose-area products (DAP) were obtained from the electronically logged protocol. Image noise was measured with an ROI. Two independent and blinded readers evaluated all images. Interreader agreement was measured with a Cohen kappa. Readers assessed overall quality and delineation of soft tissue, cortical bone, and trabecular bone on a 4-point Likert scale. RESULTS The highest DAP (3.892 ± 0.432 Gy · cm(2)) was recorded for images obtained with 60 kV and 80 mA; the lowest (0.857 ± 0.178 Gy · cm(2)) was recorded for images obtained with 50 kV and 32 mA. Noise was highest when a combination of 50 kV and 32 mA (389 ± 26.6) was used and lowest when a combination of 60 kV and 80 mA (218 ± 12.3) was used. The amount of noise on images acquired using 60 kV and 80 mA was statistically significantly different from the amount measured on all other images (p < 0.0001). Interreader agreement was excellent (κ = 0.93). Delineation of anatomy and overall quality were scored best on images obtained with 60 kV and 80 mA and worst on images obtained with 50 kV and 32 mA. The difference in delineation and quality on images obtained using 50 kV and 40 mA was not statistically significantly different compared with images obtained using 60 kV and 80 mA. CONCLUSION Significant dose reduction for tomosynthesis of the wrist is possible while image quality and delineation of anatomic structures remain preserved.

Statistics

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:20 October 2016
Deposited On:28 Oct 2016 12:01
Last Modified:31 Jan 2017 08:09
Publisher:American Roentgen Ray Society
ISSN:0361-803X
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.2214/AJR.16.16729
PubMed ID:27762599

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.