Header

UZH-Logo

Maintenance Infos

First percutaneous implantation of a completely tissue-engineered self-expanding pulmonary heart valve prosthesis using a newly developed delivery system: a feasibility study in sheep


Spriestersbach, Hendrik; Prudlo, Antonia; Bartosch, Marco; Sanders, Bart; Radtke, Torben; Baaijens, Frank P T; Hoerstrup, Simon P; Berger, Felix; Schmitt, Boris (2017). First percutaneous implantation of a completely tissue-engineered self-expanding pulmonary heart valve prosthesis using a newly developed delivery system: a feasibility study in sheep. Cardiovascular Intervention and Therapeutics, 32(1):36-47.

Abstract

In a European consortium, a decellularized tissue-engineered heart valve (dTEHV) based on vessel-derived cells, a fast-degrading scaffold and a self-expanding stent has been developed. The aim of this study was to demonstrate that percutaneous delivery is feasible. To implant this valve prosthesis transcutaneously into pulmonary position, a catheter delivery system was designed and custom made. Three sheep underwent transjugular prototype implantation. Intracardiac echocardiography (ICE), angiography and computed tomography (CT) were applied to assess the position, morphology, function and dimensions of the stented dTEHV. One animal was killed 3 h after implantation and two animals were followed up for 12 weeks. Explanted valves were analyzed macroscopically and microscopically. In all animals, the percutaneous implantation of the stented dTEHV was successful. The prototype delivery system worked at first attempt in all animals. In the first implantation a 22 F system was used: the valve was slightly damaged during crimping. Loading was difficult due to valve-catheter mismatch in volume. In the second and third implantation a 26 F system was used: the valves fitted adequately and stayed intact. Following implantation, these two valves showed moderate regurgitation due to insufficient coaptation. During follow-up, regurgitation increased due to shortened leaflets. At explantation, macroscopic and microscopic analysis confirmed the second and third valve to be intact. Histology revealed autologous recellularization of the decellularized valve after 12 weeks in vivo. It was demonstrated that completely in vitro tissue-engineered heart valves are thin and stable enough to be crimped and implanted transvenously into pulmonary position.

Abstract

In a European consortium, a decellularized tissue-engineered heart valve (dTEHV) based on vessel-derived cells, a fast-degrading scaffold and a self-expanding stent has been developed. The aim of this study was to demonstrate that percutaneous delivery is feasible. To implant this valve prosthesis transcutaneously into pulmonary position, a catheter delivery system was designed and custom made. Three sheep underwent transjugular prototype implantation. Intracardiac echocardiography (ICE), angiography and computed tomography (CT) were applied to assess the position, morphology, function and dimensions of the stented dTEHV. One animal was killed 3 h after implantation and two animals were followed up for 12 weeks. Explanted valves were analyzed macroscopically and microscopically. In all animals, the percutaneous implantation of the stented dTEHV was successful. The prototype delivery system worked at first attempt in all animals. In the first implantation a 22 F system was used: the valve was slightly damaged during crimping. Loading was difficult due to valve-catheter mismatch in volume. In the second and third implantation a 26 F system was used: the valves fitted adequately and stayed intact. Following implantation, these two valves showed moderate regurgitation due to insufficient coaptation. During follow-up, regurgitation increased due to shortened leaflets. At explantation, macroscopic and microscopic analysis confirmed the second and third valve to be intact. Histology revealed autologous recellularization of the decellularized valve after 12 weeks in vivo. It was demonstrated that completely in vitro tissue-engineered heart valves are thin and stable enough to be crimped and implanted transvenously into pulmonary position.

Statistics

Altmetrics

Downloads

0 downloads since deposited on 03 Feb 2017
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute for Regenerative Medicine (IREM)
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:January 2017
Deposited On:03 Feb 2017 09:21
Last Modified:28 Feb 2017 08:32
Publisher:Springer
ISSN:1868-4300
Publisher DOI:https://doi.org/10.1007/s12928-016-0396-y
PubMed ID:27139179

Download

Preview Icon on Download
Content: Published Version
Filetype: PDF - Registered users only
Size: 6MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations