The effect of defective PET detectors in clinical simultaneous [(18)F]FDG time-of-flight PET/MR imaging


Ter Voert, Edwin E G W; Delso, Gaspar; de Galiza Barbosa, Felipe; Huellner, Martin; Veit-Haibach, Patrick (2017). The effect of defective PET detectors in clinical simultaneous [(18)F]FDG time-of-flight PET/MR imaging. Molecular Imaging and Biology, 19(4):626-635.

Abstract

PURPOSE The purpose of this study was to evaluate the effect of defective positron emission tomography (PET) detectors on clinical PET image quality in simultaneous PET/magnetic resonance imaging (MRI) for both time-of-flight (TOF) and non-TOF reconstructed images. PROCEDURES A total of six patients with various malignant tumors were included and underwent a 2-deoxy-2-[(18)F]fluoro-D-glucose PET scan in a fully functional simultaneous TOF PET/MRI. TOF and non-TOF PET images were reconstructed before and after simulating defective detector units. All images were clinically assessed and scored. In addition, a quantitative assessment was performed. Differences were ascertained and compared using the Wilcoxon matched pairs signed-rank test. RESULTS Without TOF, the image artifacts introduced by one defective detector unit already started to degrade the overall image quality. It reduced the confidence and could lead to a change in diagnosis. Simulating three or five defective detector units resulted in more artifacts and further reduced overall image quality and confidence. By including TOF information, the effects were mitigated: Images reconstructed with one defective detector unit had similar scores as the ones without defective units. The average absolute percentage error for one, three, and five defective detector units were respectively 8, 20, and 37 % for the non-TOF cases and only 5, 11, and 19 % for the TOF cases. CONCLUSION Our study indicates that PET image artifacts due to (simulated) defective detectors are significantly mitigated with the integration of TOF information in simultaneous PET/MR. One defective detector unit introduces, on average, a 5 % absolute percentage error. However, in TOF imaging, even in cases with one or three defective units for head and neck imaging and one defective unit for chest and abdominal imaging, overall image quality, artifact scoring, and reader confidence are not significantly degraded.

Abstract

PURPOSE The purpose of this study was to evaluate the effect of defective positron emission tomography (PET) detectors on clinical PET image quality in simultaneous PET/magnetic resonance imaging (MRI) for both time-of-flight (TOF) and non-TOF reconstructed images. PROCEDURES A total of six patients with various malignant tumors were included and underwent a 2-deoxy-2-[(18)F]fluoro-D-glucose PET scan in a fully functional simultaneous TOF PET/MRI. TOF and non-TOF PET images were reconstructed before and after simulating defective detector units. All images were clinically assessed and scored. In addition, a quantitative assessment was performed. Differences were ascertained and compared using the Wilcoxon matched pairs signed-rank test. RESULTS Without TOF, the image artifacts introduced by one defective detector unit already started to degrade the overall image quality. It reduced the confidence and could lead to a change in diagnosis. Simulating three or five defective detector units resulted in more artifacts and further reduced overall image quality and confidence. By including TOF information, the effects were mitigated: Images reconstructed with one defective detector unit had similar scores as the ones without defective units. The average absolute percentage error for one, three, and five defective detector units were respectively 8, 20, and 37 % for the non-TOF cases and only 5, 11, and 19 % for the TOF cases. CONCLUSION Our study indicates that PET image artifacts due to (simulated) defective detectors are significantly mitigated with the integration of TOF information in simultaneous PET/MR. One defective detector unit introduces, on average, a 5 % absolute percentage error. However, in TOF imaging, even in cases with one or three defective units for head and neck imaging and one defective unit for chest and abdominal imaging, overall image quality, artifact scoring, and reader confidence are not significantly degraded.

Statistics

Citations

Dimensions.ai Metrics
1 citation in Web of Science®
1 citation in Scopus®
1 citation in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 21 Nov 2016
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2017
Deposited On:21 Nov 2016 10:34
Last Modified:19 Feb 2018 07:25
Publisher:Springer
ISSN:1536-1632
OA Status:Closed
Publisher DOI:https://doi.org/10.1007/s11307-016-1023-0
PubMed ID:27812923

Download