Single-shot spiral imaging enabled by an expanded encoding model: demonstration in diffusion MRI


Wilm, B; Barmet, C; Gross, S; Kasper, L; Vannesjo, S J; Haeberlin, M; Dietrich, B E; Brunner, D O; Schmid, Thomas; Pruessmann, K P (2017). Single-shot spiral imaging enabled by an expanded encoding model: demonstration in diffusion MRI. Magnetic Resonance in Medicine, 77:83-91.

Abstract

Purpose The purpose of this work was to improve the quality of single-shot spiral MRI and demonstrate its application for diffusion-weighted imaging.
Methods Image formation is based on an expanded encoding model that accounts for dynamic magnetic fields up to third order in space, nonuniform static B0, and coil sensitivity encoding. The encoding model is determined by B0 mapping, sensitivity mapping, and concurrent field monitoring. Reconstruction is performed by iterative inversion of the expanded signal equations. Diffusion-tensor imaging with single-shot spiral readouts is performed in a phantom and in vivo, using a clinical 3T instrument. Image quality is assessed in terms of artefact levels, image congruence, and the influence of the different encoding factors.
Results Using the full encoding model, diffusion-weighted single-shot spiral imaging of high quality is accomplished both in vitro and in vivo. Accounting for actual field dynamics, including higher orders, is found to be critical to suppress blurring, aliasing, and distortion. Enhanced image congruence permitted data fusion and diffusion tensor analysis without coregistration. Conclusion Use of an expanded signal model largely overcomes the traditional vulnerability of spiral imaging with long readouts. It renders single-shot spirals competitive with echo-planar readouts and thus deploys shorter echo times and superior readout efficiency for diffusion imaging and further prospective applications.

Abstract

Purpose The purpose of this work was to improve the quality of single-shot spiral MRI and demonstrate its application for diffusion-weighted imaging.
Methods Image formation is based on an expanded encoding model that accounts for dynamic magnetic fields up to third order in space, nonuniform static B0, and coil sensitivity encoding. The encoding model is determined by B0 mapping, sensitivity mapping, and concurrent field monitoring. Reconstruction is performed by iterative inversion of the expanded signal equations. Diffusion-tensor imaging with single-shot spiral readouts is performed in a phantom and in vivo, using a clinical 3T instrument. Image quality is assessed in terms of artefact levels, image congruence, and the influence of the different encoding factors.
Results Using the full encoding model, diffusion-weighted single-shot spiral imaging of high quality is accomplished both in vitro and in vivo. Accounting for actual field dynamics, including higher orders, is found to be critical to suppress blurring, aliasing, and distortion. Enhanced image congruence permitted data fusion and diffusion tensor analysis without coregistration. Conclusion Use of an expanded signal model largely overcomes the traditional vulnerability of spiral imaging with long readouts. It renders single-shot spirals competitive with echo-planar readouts and thus deploys shorter echo times and superior readout efficiency for diffusion imaging and further prospective applications.

Statistics

Altmetrics

Downloads

1 download since deposited on 03 Feb 2017
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Language:English
Date:2017
Deposited On:03 Feb 2017 15:37
Last Modified:03 Feb 2017 15:37
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:0740-3194
Publisher DOI:https://doi.org/10.1002/mrm.26493
PubMed ID:27770473

Download