Head-to-head comparison of adaptive statistical and model-based iterative reconstruction algorithms for submillisievert coronary CT angiography


Benz, Dominik C; Fuchs, Tobias A; Gräni, Christoph; Studer Bruengger, Annina A; Clerc, Olivier F; Mikulicic, Fran; Messerli, Michael; Stehli, Julia; Possner, Mathias; Pazhenkottil, Aju P; Gaemperli, Oliver; Kaufmann, Philipp A; Buechel, Ronny R (2017). Head-to-head comparison of adaptive statistical and model-based iterative reconstruction algorithms for submillisievert coronary CT angiography. European Heart Journal. Cardiovascular Imaging:Epub ahead of print.

Abstract

Aims: Iterative reconstruction (IR) algorithms allow for a significant reduction in radiation dose of coronary computed tomography angiography (CCTA). We performed a head-to-head comparison of adaptive statistical IR (ASiR) and model-based IR (MBIR) algorithms to assess their impact on quantitative image parameters and diagnostic accuracy for submillisievert CCTA.
Methods and results: CCTA datasets of 91 patients were reconstructed using filtered back projection (FBP), increasing contributions of ASiR (20, 40, 60, 80, and 100%), and MBIR. Signal and noise were measured in the aortic root to calculate signal-to-noise ratio (SNR). In a subgroup of 36 patients, diagnostic accuracy of ASiR 40%, ASiR 100%, and MBIR for diagnosis of coronary artery disease (CAD) was compared with invasive coronary angiography. Median radiation dose was 0.21 mSv for CCTA. While increasing levels of ASiR gradually reduced image noise compared with FBP (up to - 48%, P < 0.001), MBIR provided largest noise reduction (-79% compared with FBP) outperforming ASiR (-59% compared with ASiR 100%; P < 0.001). Increased noise and lower SNR with ASiR 40% and ASiR 100% resulted in substantially lower diagnostic accuracy to detect CAD as diagnosed by invasive coronary angiography compared with MBIR: sensitivity and specificity were 100 and 37%, 100 and 57%, and 100 and 74% for ASiR 40%, ASiR 100%, and MBIR, respectively.
Conclusion: MBIR offers substantial noise reduction with increased SNR, paving the way for implementation of submillisievert CCTA protocols in clinical routine. In contrast, inferior noise reduction by ASiR negatively affects diagnostic accuracy of submillisievert CCTA for CAD detection.

Abstract

Aims: Iterative reconstruction (IR) algorithms allow for a significant reduction in radiation dose of coronary computed tomography angiography (CCTA). We performed a head-to-head comparison of adaptive statistical IR (ASiR) and model-based IR (MBIR) algorithms to assess their impact on quantitative image parameters and diagnostic accuracy for submillisievert CCTA.
Methods and results: CCTA datasets of 91 patients were reconstructed using filtered back projection (FBP), increasing contributions of ASiR (20, 40, 60, 80, and 100%), and MBIR. Signal and noise were measured in the aortic root to calculate signal-to-noise ratio (SNR). In a subgroup of 36 patients, diagnostic accuracy of ASiR 40%, ASiR 100%, and MBIR for diagnosis of coronary artery disease (CAD) was compared with invasive coronary angiography. Median radiation dose was 0.21 mSv for CCTA. While increasing levels of ASiR gradually reduced image noise compared with FBP (up to - 48%, P < 0.001), MBIR provided largest noise reduction (-79% compared with FBP) outperforming ASiR (-59% compared with ASiR 100%; P < 0.001). Increased noise and lower SNR with ASiR 40% and ASiR 100% resulted in substantially lower diagnostic accuracy to detect CAD as diagnosed by invasive coronary angiography compared with MBIR: sensitivity and specificity were 100 and 37%, 100 and 57%, and 100 and 74% for ASiR 40%, ASiR 100%, and MBIR, respectively.
Conclusion: MBIR offers substantial noise reduction with increased SNR, paving the way for implementation of submillisievert CCTA protocols in clinical routine. In contrast, inferior noise reduction by ASiR negatively affects diagnostic accuracy of submillisievert CCTA for CAD detection.

Statistics

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:15 February 2017
Deposited On:17 May 2017 09:09
Last Modified:09 Dec 2017 00:58
Publisher:Oxford University Press
ISSN:2047-2404
Publisher DOI:https://doi.org/10.1093/ehjci/jex008
PubMed ID:28200212

Download

Full text not available from this repository.
View at publisher