Prototype metal artefact reduction algorithm in flat panel computed tomography - evaluation in patients undergoing transarterial hepatic radioembolisation


Hamie, Qeumars Mustafa; Kobe, Adrian Raoul; Mietzsch, Leif; Manhart, Michael; Puippe, Gilbert Dominique; Pfammatter, Thomas; Guggenberger, Roman (2017). Prototype metal artefact reduction algorithm in flat panel computed tomography - evaluation in patients undergoing transarterial hepatic radioembolisation. European Radiology:Epub ahead of print.

Abstract

OBJECTIVE: To investigate the effect of an on-site prototype metal artefact reduction (MAR) algorithm in cone-beam CT-catheter-arteriography (CBCT-CA) in patients undergoing transarterial radioembolisation (RE) of hepatic masses.
METHODS AND MATERIALS: Ethical board approved retrospective study of 29 patients (mean 63.7±13.7 years, 11 female), including 16 patients with arterial metallic coils, undergoing CBCT-CA (8s scan, 200 degrees rotation, 397 projections). Image reconstructions with and without prototype MAR algorithm were evaluated quantitatively (streak-artefact attenuation changes) and qualitatively (visibility of hepatic parenchyma and vessels) in near- (<1cm) and far-field (>3cm) of artefact sources (metallic coils and catheters). Quantitative and qualitative measurements of uncorrected and MAR corrected images and different artefact sources were compared RESULTS: Quantitative evaluation showed significant reduction of near- and far-field streak-artefacts with MAR for both artefact sources (p<0.001), while remaining stable for unaffected organs (all p>0.05). Inhomogeneities of attenuation values were significantly higher for metallic coils compared to catheters (p<0.001) and decreased significantly for both after MAR (p<0.001). Qualitative image scores were significantly improved after MAR (all p<0.003) with by trend higher artefact degrees for metallic coils compared to catheters.
CONCLUSION: In patients undergoing CBCT-CA for transarterial RE, prototype MAR algorithm improves image quality in proximity of metallic coil and catheter artefacts.
KEY POINTS: Metal objects cause artefacts in cone-beam computed tomography (CBCT) imaging. • These artefacts can be corrected by metal artefact reduction (MAR) algorithms. • Corrected images show significantly better visibility of nearby hepatic vessels and tissue. • Better visibility may facilitate image interpretation, save time and radiation exposure.

Abstract

OBJECTIVE: To investigate the effect of an on-site prototype metal artefact reduction (MAR) algorithm in cone-beam CT-catheter-arteriography (CBCT-CA) in patients undergoing transarterial radioembolisation (RE) of hepatic masses.
METHODS AND MATERIALS: Ethical board approved retrospective study of 29 patients (mean 63.7±13.7 years, 11 female), including 16 patients with arterial metallic coils, undergoing CBCT-CA (8s scan, 200 degrees rotation, 397 projections). Image reconstructions with and without prototype MAR algorithm were evaluated quantitatively (streak-artefact attenuation changes) and qualitatively (visibility of hepatic parenchyma and vessels) in near- (<1cm) and far-field (>3cm) of artefact sources (metallic coils and catheters). Quantitative and qualitative measurements of uncorrected and MAR corrected images and different artefact sources were compared RESULTS: Quantitative evaluation showed significant reduction of near- and far-field streak-artefacts with MAR for both artefact sources (p<0.001), while remaining stable for unaffected organs (all p>0.05). Inhomogeneities of attenuation values were significantly higher for metallic coils compared to catheters (p<0.001) and decreased significantly for both after MAR (p<0.001). Qualitative image scores were significantly improved after MAR (all p<0.003) with by trend higher artefact degrees for metallic coils compared to catheters.
CONCLUSION: In patients undergoing CBCT-CA for transarterial RE, prototype MAR algorithm improves image quality in proximity of metallic coil and catheter artefacts.
KEY POINTS: Metal objects cause artefacts in cone-beam computed tomography (CBCT) imaging. • These artefacts can be corrected by metal artefact reduction (MAR) algorithms. • Corrected images show significantly better visibility of nearby hepatic vessels and tissue. • Better visibility may facilitate image interpretation, save time and radiation exposure.

Statistics

Altmetrics

Downloads

0 downloads since deposited on 30 Aug 2017
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Uncontrolled Keywords:Abdomen; Cone-beam computed tomography; Liver; Metal artefact reduction; Transarterial radioembolisation
Language:English
Date:14 July 2017
Deposited On:30 Aug 2017 16:38
Last Modified:31 Aug 2017 07:30
Publisher:Springer
ISSN:0938-7994
Publisher DOI:https://doi.org/10.1007/s00330-017-4946-1
PubMed ID:28710583

Download

Preview Icon on Download
Content: Published Version
Language: English
Filetype: PDF - Registered users only
Size: 1MB
View at publisher