Ultrasound needle visibility in contrast mode imaging: an in vitro and ex vivo study


Rominger, Marga B; Martini, Katharina; Dappa, Evelyn; Puippe, Gilbert Dominique; Klingmüller, Volker; Frauenfelder, Thomas; Sanabria, Sergio J (2017). Ultrasound needle visibility in contrast mode imaging: an in vitro and ex vivo study. Ultrasound international open, 3(2):E82-E88.

Abstract

PURPOSE: To evaluate needle visibility in ultrasound under contrast mode conditions.
MATERIALS AND METHODS: Needle visibility was evaluated for bevel, EchoTip (®) and shaft of 18G Chiba biopsy needle with a 9 MHz linear probe (GE Logiq E9). Insertion angles varied between 30°(steep) and 90°(parallel to the probe surface). The acoustic output varied from 5-28%. 2 different contrast mode presets with either 'Amplitude Modulation' (Penetration) or 'Phase Inversion Harmonics' (High Resolution) were assessed. All other imaging parameters were kept constant. The visibility of bevel, EchoTip (®) and shaft was assessed for grayscale and color-coded images with a 3-point Likert-like scale (not, poorly, well visible) by 2 independent readers. The echogenicity of the needle bevel, EchoTip (®) and shaft was assessed in deciBel (dB) on the color-coded images.
RESULTS: With the parallel insertion angle, all needle areas were well visible. With steep insertion the EchoTip (®) was the only visible area. High Resolution was superior to Penetration (p<0.001). The visibility and echogenicity of the needle bevel ( rgrayscale =0.109, p grayscale =0.178; rcolor-coded =0.236, p color-coded =0.266; rdB =0.956, p dB =0.001), EchoTip (®) ( rgrayscale += 0.477, p grayscale += 0.018; rcolor-coded =0.540, p color-coded += 0.006; rdB =0.911, p dB =0.001) and shaft ( rgrayscale =0.124, p grayscale =0.563; rcolor-coded =0.061, p color-coded += 0.775; rdB += 0.926, p dB =0.001) increased with increasing acoustic output. Grayscale images were superior to color-coded images for needle visibility (p=0.004).
CONCLUSION: Parallel needle insertion, use of an echogenic tip, adequate choice of presets, increased acoustic output, and dual view of grayscale and color-coded images improve needle visibility in ultrasound under contrast mode conditions.

Abstract

PURPOSE: To evaluate needle visibility in ultrasound under contrast mode conditions.
MATERIALS AND METHODS: Needle visibility was evaluated for bevel, EchoTip (®) and shaft of 18G Chiba biopsy needle with a 9 MHz linear probe (GE Logiq E9). Insertion angles varied between 30°(steep) and 90°(parallel to the probe surface). The acoustic output varied from 5-28%. 2 different contrast mode presets with either 'Amplitude Modulation' (Penetration) or 'Phase Inversion Harmonics' (High Resolution) were assessed. All other imaging parameters were kept constant. The visibility of bevel, EchoTip (®) and shaft was assessed for grayscale and color-coded images with a 3-point Likert-like scale (not, poorly, well visible) by 2 independent readers. The echogenicity of the needle bevel, EchoTip (®) and shaft was assessed in deciBel (dB) on the color-coded images.
RESULTS: With the parallel insertion angle, all needle areas were well visible. With steep insertion the EchoTip (®) was the only visible area. High Resolution was superior to Penetration (p<0.001). The visibility and echogenicity of the needle bevel ( rgrayscale =0.109, p grayscale =0.178; rcolor-coded =0.236, p color-coded =0.266; rdB =0.956, p dB =0.001), EchoTip (®) ( rgrayscale += 0.477, p grayscale += 0.018; rcolor-coded =0.540, p color-coded += 0.006; rdB =0.911, p dB =0.001) and shaft ( rgrayscale =0.124, p grayscale =0.563; rcolor-coded =0.061, p color-coded += 0.775; rdB += 0.926, p dB =0.001) increased with increasing acoustic output. Grayscale images were superior to color-coded images for needle visibility (p=0.004).
CONCLUSION: Parallel needle insertion, use of an echogenic tip, adequate choice of presets, increased acoustic output, and dual view of grayscale and color-coded images improve needle visibility in ultrasound under contrast mode conditions.

Statistics

Altmetrics

Downloads

6 downloads since deposited on 30 Aug 2017
6 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:April 2017
Deposited On:30 Aug 2017 17:19
Last Modified:24 Sep 2017 05:03
Publisher:Georg Thieme Verlag
ISSN:2199-7152
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1055/s-0043-101511
PubMed ID:28653045

Download

Download PDF  'Ultrasound needle visibility in contrast mode imaging: an in vitro and ex vivo study'.
Preview
Content: Published Version
Language: English
Filetype: PDF
Size: 880kB
View at publisher
Licence: Creative Commons: Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)