Automatic valve plane localization in myocardial perfusion SPECT/CT by machine learning: anatomic and clinical validation


Betancur, Julian; Rubeaux, Mathieu; Fuchs, Tobias A; Otaki, Yuka; Arnson, Yoav; Slipczuk, Leandro; Benz, Dominik C; Germano, Guido; Dey, Damini; Lin, Chih-Jen; Berman, Daniel S; Kaufmann, Philipp A; Slomka, Piotr J (2017). Automatic valve plane localization in myocardial perfusion SPECT/CT by machine learning: anatomic and clinical validation. Journal of Nuclear Medicine, 58(6):961-967.

Abstract

Precise definition of the mitral valve plane (VP) during segmentation of the left ventricle for SPECT myocardial perfusion imaging (MPI) quantification often requires manual adjustment, which affects the quantification of perfusion. We developed a machine learning approach using support vector machines (SVM) for automatic VP placement. Methods: A total of 392 consecutive patients undergoing 99mTc-tetrofosmin stress (5 min; mean ± SD, 350 ± 54 MBq) and rest (5 min; 1,024 ± 153 MBq) fast SPECT MPI attenuation corrected (AC) by CT and same-day coronary CT angiography were studied; included in the 392 patients were 48 patients who underwent invasive coronary angiography and had no known coronary artery disease. The left ventricle was segmented with standard clinical software (quantitative perfusion SPECT) by 2 experts, adjusting the VP if needed. Two-class SVM models were computed from the expert placements with 10-fold cross validation to separate the patients used for training and those used for validation. SVM probability estimates were used to compute the best VP position. Automatic VP localizations on AC and non-AC images were compared with expert placement on coronary CT angiography. Stress and rest total perfusion deficits and detection of per-vessel obstructive stenosis by invasive coronary angiography were also compared. Results: Bland-Altman 95% confidence intervals (CIs) for VP localization by SVM and experts for AC stress images (bias, 1; 95% CI, -5 to 7 mm) and AC rest images (bias, 1; 95% CI, -7 to 10 mm) were narrower than interexpert 95% CIs for AC stress images (bias, 0; 95% CI, -8 to 8 mm) and AC rest images (bias, 0; 95% CI, -10 to 10 mm) (P < 0.01). Bland-Altman 95% CIs for VP localization by SVM and experts for non-AC stress images (bias, 1; 95% CI, -4 to 6 mm) and non-AC rest images (bias, 2; 95% CI, -7 to 10 mm) were similar to interexpert 95% CIs for non-AC stress images (bias, 0; 95% CI, -6 to 5 mm) and non-AC rest images (bias, -1; 95% CI, -9 to 7 mm) (P was not significant [NS]). For regional detection of obstructive stenosis, ischemic total perfusion deficit areas under the receiver operating characteristic curve for the 2 experts (AUC, 0.79 [95% CI, 0.7-0.87]; AUC, 0.81 [95% CI, 0.73-0.89]) and the SVM (0.82 [0.74-0.9]) for AC data were the same (P = NS) and were higher than those for the unadjusted VP (0.63 [0.53-0.73]) (P < 0.01). Similarly, for non-AC data, areas under the receiver operating characteristic curve for the experts (AUC, 0.77 [95% CI, 0.69-0.89]; AUC, 0.8 [95% CI, 0.72-0.88]) and the SVM (0.79 [0.71-0.87]) were the same (P = NS) and were higher than those for the unadjusted VP (0.65 [0.56-0.75]) (P < 0.01). Conclusion: Machine learning with SVM allows automatic and accurate VP localization, decreasing user dependence in SPECT MPI quantification.

Abstract

Precise definition of the mitral valve plane (VP) during segmentation of the left ventricle for SPECT myocardial perfusion imaging (MPI) quantification often requires manual adjustment, which affects the quantification of perfusion. We developed a machine learning approach using support vector machines (SVM) for automatic VP placement. Methods: A total of 392 consecutive patients undergoing 99mTc-tetrofosmin stress (5 min; mean ± SD, 350 ± 54 MBq) and rest (5 min; 1,024 ± 153 MBq) fast SPECT MPI attenuation corrected (AC) by CT and same-day coronary CT angiography were studied; included in the 392 patients were 48 patients who underwent invasive coronary angiography and had no known coronary artery disease. The left ventricle was segmented with standard clinical software (quantitative perfusion SPECT) by 2 experts, adjusting the VP if needed. Two-class SVM models were computed from the expert placements with 10-fold cross validation to separate the patients used for training and those used for validation. SVM probability estimates were used to compute the best VP position. Automatic VP localizations on AC and non-AC images were compared with expert placement on coronary CT angiography. Stress and rest total perfusion deficits and detection of per-vessel obstructive stenosis by invasive coronary angiography were also compared. Results: Bland-Altman 95% confidence intervals (CIs) for VP localization by SVM and experts for AC stress images (bias, 1; 95% CI, -5 to 7 mm) and AC rest images (bias, 1; 95% CI, -7 to 10 mm) were narrower than interexpert 95% CIs for AC stress images (bias, 0; 95% CI, -8 to 8 mm) and AC rest images (bias, 0; 95% CI, -10 to 10 mm) (P < 0.01). Bland-Altman 95% CIs for VP localization by SVM and experts for non-AC stress images (bias, 1; 95% CI, -4 to 6 mm) and non-AC rest images (bias, 2; 95% CI, -7 to 10 mm) were similar to interexpert 95% CIs for non-AC stress images (bias, 0; 95% CI, -6 to 5 mm) and non-AC rest images (bias, -1; 95% CI, -9 to 7 mm) (P was not significant [NS]). For regional detection of obstructive stenosis, ischemic total perfusion deficit areas under the receiver operating characteristic curve for the 2 experts (AUC, 0.79 [95% CI, 0.7-0.87]; AUC, 0.81 [95% CI, 0.73-0.89]) and the SVM (0.82 [0.74-0.9]) for AC data were the same (P = NS) and were higher than those for the unadjusted VP (0.63 [0.53-0.73]) (P < 0.01). Similarly, for non-AC data, areas under the receiver operating characteristic curve for the experts (AUC, 0.77 [95% CI, 0.69-0.89]; AUC, 0.8 [95% CI, 0.72-0.88]) and the SVM (0.79 [0.71-0.87]) were the same (P = NS) and were higher than those for the unadjusted VP (0.65 [0.56-0.75]) (P < 0.01). Conclusion: Machine learning with SVM allows automatic and accurate VP localization, decreasing user dependence in SPECT MPI quantification.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

1 download since deposited on 08 Feb 2018
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:June 2017
Deposited On:08 Feb 2018 20:31
Last Modified:01 Mar 2018 01:56
Publisher:Society of Nuclear Medicine
ISSN:0161-5505
OA Status:Green
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.2967/jnumed.116.179911
PubMed ID:27811121

Download

Download PDF  'Automatic valve plane localization in myocardial perfusion SPECT/CT by machine learning: anatomic and clinical validation'.
Preview
Content: Accepted Version
Filetype: PDF
Size: 2MB
View at publisher