Are Commercially-available Precontoured Anatomical Clavicle Plating Systems Offering the Purported Superior Optimum Fitting to the Clavicle? A Cadaveric Analysis and Review of Literature


Bauer, D E; Hingsammer, A; Schenk, P; Vlachopoulos, L; Imam, M A; Fürnstahl, P; Meyer, D C (2018). Are Commercially-available Precontoured Anatomical Clavicle Plating Systems Offering the Purported Superior Optimum Fitting to the Clavicle? A Cadaveric Analysis and Review of Literature. Orthopaedics & Traumatology, Surgery & Research (OTSR):1877-0568.

Abstract

PURPOSE The indication for operative treatment of displaced midshaft clavicle fractures remains controversial. However, if plate fixation is considered, implant prominence and skin irritation are the most common causes for re-operation. Low profile implants as well as closely contouring plates to the individual anatomy may reduce these complications. The aim of this study was to compare the fitting accuracy and implant prominence of 3.5mm pelvic reconstruction plates (PRP) with pre-contoured anatomical clavicle plates (PACP) for midshaft clavicle fractures.
METHODS Three-dimensional data of the largest, median and smallest male and female clavicle of an existing database of 89 cadaveric clavicles were included for analysis. A three-dimensional model of a commercially available PACP was used for digitally positioning of the plate on the segmented clavicles. Three-dimensional printouts of each clavicle were produced and the 3.5mm reconstruction plates were manually bent and positioned by the senior author. Computed tomography scans and three-dimensional reconstructions were then obtained to digitally compare the fitting accuracy and implant prominence.
RESULTS Pelvic reconstruction plates offered superior fitting accuracy and lower implant prominence compared to PACP. The largest difference in implant prominence was observed in large sized female clavicles and measured 3.6mm.
CONCLUSION Both, the less costly PRP plates and commercially available PACP for midshaft fractures of the clavicle demonstrated a clinically acceptable fitting accuracy. The manually bent pelvic-reconstruction plates demonstrated reduced implant prominence with superior fitting. Hypothetically this might contribute to a reduced rate of reoperation.

Abstract

PURPOSE The indication for operative treatment of displaced midshaft clavicle fractures remains controversial. However, if plate fixation is considered, implant prominence and skin irritation are the most common causes for re-operation. Low profile implants as well as closely contouring plates to the individual anatomy may reduce these complications. The aim of this study was to compare the fitting accuracy and implant prominence of 3.5mm pelvic reconstruction plates (PRP) with pre-contoured anatomical clavicle plates (PACP) for midshaft clavicle fractures.
METHODS Three-dimensional data of the largest, median and smallest male and female clavicle of an existing database of 89 cadaveric clavicles were included for analysis. A three-dimensional model of a commercially available PACP was used for digitally positioning of the plate on the segmented clavicles. Three-dimensional printouts of each clavicle were produced and the 3.5mm reconstruction plates were manually bent and positioned by the senior author. Computed tomography scans and three-dimensional reconstructions were then obtained to digitally compare the fitting accuracy and implant prominence.
RESULTS Pelvic reconstruction plates offered superior fitting accuracy and lower implant prominence compared to PACP. The largest difference in implant prominence was observed in large sized female clavicles and measured 3.6mm.
CONCLUSION Both, the less costly PRP plates and commercially available PACP for midshaft fractures of the clavicle demonstrated a clinically acceptable fitting accuracy. The manually bent pelvic-reconstruction plates demonstrated reduced implant prominence with superior fitting. Hypothetically this might contribute to a reduced rate of reoperation.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

1 download since deposited on 27 Mar 2018
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Balgrist University Hospital, Swiss Spinal Cord Injury Center
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:16 March 2018
Deposited On:27 Mar 2018 09:44
Last Modified:13 Apr 2018 11:48
Publisher:Elsevier
ISSN:1877-0568
OA Status:Closed
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1016/j.otsr.2018.01.013
PubMed ID:29555558

Download