Coronary artery bypass grafts: ECG-gated multi-detector row CT angiography--influence of image reconstruction interval on graft visibility


Willmann, J K; Weishaupt, D; Kobza, R; Verdun, F R; Seifert, Burkhardt; Marincek, B; Boehm, T (2004). Coronary artery bypass grafts: ECG-gated multi-detector row CT angiography--influence of image reconstruction interval on graft visibility. Radiology, 232(2):568-577.

Abstract

PURPOSE: To evaluate the influence of different reconstruction intervals of retrospectively electrocardiographically (ECG)-gated multi-detector row computed tomographic (CT) angiography on image quality of different segments of various types of coronary artery bypass grafts. MATERIALS AND METHODS: Twenty consecutive patients with 62 grafts underwent retrospectively ECG-gated four-channel multi-detector row CT angiography and conventional coronary angiography. Raw helical CT data were reconstructed at 0%-90% of the cardiac cycle in increments of 10%. Each graft was separated into three segments (proximal segment, graft body, and distal anastomosis). Three graft types were identified according to site of distal anastomosis. Two readers assessed image quality of segments and graft types. Effective radiation dose was calculated. RESULTS: Best image quality of all segments was obtained at a reconstruction interval of 50%-70% of the cardiac cycle. Image quality of the proximal segment did not vary significantly with different reconstruction intervals (analysis of variance, P =.8), whereas image quality of the graft body and distal anastomosis changed significantly with varying reconstruction intervals (P <.001). Distal anastomosis and body of types 1 and 2 grafts were best seen at 60%-70% of the cardiac cycle, whereas distal anastomosis and body of type 3 grafts were best visualized at 50%. Accuracy of CT angiography for detection of graft patency was 94% for reader 1 and 95% for reader 2. Effective dose for CT was 11.4 mSv for both men and women. Mean effective dose for angiography was 2.1 mSv for men and women. CONCLUSION: Optimal selection of reconstruction interval improves image quality of the graft body and of distal anastomosis in particular.

Abstract

PURPOSE: To evaluate the influence of different reconstruction intervals of retrospectively electrocardiographically (ECG)-gated multi-detector row computed tomographic (CT) angiography on image quality of different segments of various types of coronary artery bypass grafts. MATERIALS AND METHODS: Twenty consecutive patients with 62 grafts underwent retrospectively ECG-gated four-channel multi-detector row CT angiography and conventional coronary angiography. Raw helical CT data were reconstructed at 0%-90% of the cardiac cycle in increments of 10%. Each graft was separated into three segments (proximal segment, graft body, and distal anastomosis). Three graft types were identified according to site of distal anastomosis. Two readers assessed image quality of segments and graft types. Effective radiation dose was calculated. RESULTS: Best image quality of all segments was obtained at a reconstruction interval of 50%-70% of the cardiac cycle. Image quality of the proximal segment did not vary significantly with different reconstruction intervals (analysis of variance, P =.8), whereas image quality of the graft body and distal anastomosis changed significantly with varying reconstruction intervals (P <.001). Distal anastomosis and body of types 1 and 2 grafts were best seen at 60%-70% of the cardiac cycle, whereas distal anastomosis and body of type 3 grafts were best visualized at 50%. Accuracy of CT angiography for detection of graft patency was 94% for reader 1 and 95% for reader 2. Effective dose for CT was 11.4 mSv for both men and women. Mean effective dose for angiography was 2.1 mSv for men and women. CONCLUSION: Optimal selection of reconstruction interval improves image quality of the graft body and of distal anastomosis in particular.

Statistics

Citations

Dimensions.ai Metrics
27 citations in Web of Science®
32 citations in Scopus®
61 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

116 downloads since deposited on 15 Jun 2009
23 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Epidemiology, Biostatistics and Prevention Institute (EBPI)
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2004
Deposited On:15 Jun 2009 15:57
Last Modified:19 Feb 2018 20:31
Publisher:Radiological Society of North America
ISSN:0033-8419
Additional Information:Free full text article
OA Status:Green
Publisher DOI:https://doi.org/10.1148/radiol.2322030788
PubMed ID:15215552

Download

Download PDF  'Coronary artery bypass grafts: ECG-gated multi-detector row CT angiography--influence of image reconstruction interval on graft visibility'.
Preview
Filetype: PDF
Size: 1MB
View at publisher