Expiratory automatic endotracheal tube compensation reduces dynamic hyperinflation in a physical lung model


Haberthür, C; Mehlig, A; Stover, J F; Schumann, S; Möller, K; Priebe, H J; Guttmann, J (2009). Expiratory automatic endotracheal tube compensation reduces dynamic hyperinflation in a physical lung model. Critical Care, 13(1):R4.

Abstract

INTRODUCTION: The effect of expiratory endotracheal tube (ETT) resistance on dynamic lung inflation is unknown. We hypothesized that ETT resistance causes dynamic lung hyperinflation by impeding lung emptying. We further hypothesized that compensation for expiratory ETT resistance by automatic tube compensation (ATC) attenuates dynamic lung hyperinflation. METHODS: A ventilator equipped with the original ATC mode and operating in a pressure-targeted mode was connected to a physical lung model that consists of four equally sized glass bottles filled with copper wool. Inspiratory pressure, peak expiratory flow, trapped lung volume and intrinsic positive end-expiratory pressure (PEEP) were assessed at combinations of four inner ETT diameters (7.0, 7.5, 8.0 and 8.5 mm), four respiratory rates (15, 20, 25 and 30/minute), three inspiratory pressures (3.0, 4.5 and 6.0 cmH2O) and four lung compliances (113, 86, 58 and 28 ml/cmH2O). Intrinsic PEEP was measured at the end of an expiratory hold manoeuvre. RESULTS: At a given test lung compliance, inspiratory pressure and ETT size, increasing respiratory rates from 15 to 30/minutes had the following effects: inspiratory tidal volume and peak expiratory flow were decreased by means of 25% (range 0% to 51%) and 11% (8% to 12%), respectively; and trapped lung volume and intrinsic PEEP were increased by means of 25% (0% to 51%) and 26% (5% to 45%), respectively (all P < 0.025). At otherwise identical baseline conditions, introduction of expiratory ATC significantly attenuated (P < 0.025), by approximately 50%, the respiratory rate-dependent decreases in inspiratory tidal volume and the increases in trapped lung volume and intrinsic PEEP. CONCLUSIONS: In a lung model of pressure-targeted ventilation, expiratory ETT resistance caused dynamic lung hyperinflation during increases in respiratory rates, thereby reducing inspiratory tidal volume. Expiratory ATC attenuated these adverse effects.

Abstract

INTRODUCTION: The effect of expiratory endotracheal tube (ETT) resistance on dynamic lung inflation is unknown. We hypothesized that ETT resistance causes dynamic lung hyperinflation by impeding lung emptying. We further hypothesized that compensation for expiratory ETT resistance by automatic tube compensation (ATC) attenuates dynamic lung hyperinflation. METHODS: A ventilator equipped with the original ATC mode and operating in a pressure-targeted mode was connected to a physical lung model that consists of four equally sized glass bottles filled with copper wool. Inspiratory pressure, peak expiratory flow, trapped lung volume and intrinsic positive end-expiratory pressure (PEEP) were assessed at combinations of four inner ETT diameters (7.0, 7.5, 8.0 and 8.5 mm), four respiratory rates (15, 20, 25 and 30/minute), three inspiratory pressures (3.0, 4.5 and 6.0 cmH2O) and four lung compliances (113, 86, 58 and 28 ml/cmH2O). Intrinsic PEEP was measured at the end of an expiratory hold manoeuvre. RESULTS: At a given test lung compliance, inspiratory pressure and ETT size, increasing respiratory rates from 15 to 30/minutes had the following effects: inspiratory tidal volume and peak expiratory flow were decreased by means of 25% (range 0% to 51%) and 11% (8% to 12%), respectively; and trapped lung volume and intrinsic PEEP were increased by means of 25% (0% to 51%) and 26% (5% to 45%), respectively (all P < 0.025). At otherwise identical baseline conditions, introduction of expiratory ATC significantly attenuated (P < 0.025), by approximately 50%, the respiratory rate-dependent decreases in inspiratory tidal volume and the increases in trapped lung volume and intrinsic PEEP. CONCLUSIONS: In a lung model of pressure-targeted ventilation, expiratory ETT resistance caused dynamic lung hyperinflation during increases in respiratory rates, thereby reducing inspiratory tidal volume. Expiratory ATC attenuated these adverse effects.

Statistics

Citations

8 citations in Web of Science®
7 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

65 downloads since deposited on 12 Feb 2010
16 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Intensive Care Medicine
04 Faculty of Medicine > University Hospital Zurich > Division of Surgical Research
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2009
Deposited On:12 Feb 2010 07:42
Last Modified:07 Dec 2017 00:18
Publisher:BioMed Central
ISSN:1364-8535
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.1186/cc7693
PubMed ID:19166607

Download

Download PDF  'Expiratory automatic endotracheal tube compensation reduces dynamic hyperinflation in a physical lung model'.
Preview
Content: Published Version
Filetype: PDF
Size: 1MB
View at publisher
Licence: Creative Commons: Attribution 2.0 Generic (CC BY 2.0)