Evaluation of a novel haematology analyser for use with feline blood


Weissenbacher, S; Riond, B; Hofmann-Lehmann, R; Lutz, H (2011). Evaluation of a novel haematology analyser for use with feline blood. Veterinary Journal, 187(3):381-387.

Abstract

A novel haematology analyser was evaluated for its use with feline samples. Complete blood cell counts, a five-part differential count, and reticulocyte numbers were determined, and the results compared with reference data. Coefficients of correlation, Passing-Bablok regression analysis and Bland-Altmann difference plots with biases and 95% limits of agreement are reported. Precision and linearity were also studied. The instrument demonstrated very low imprecision, and the tested range of linearity exceeded the reference ranges provided by the manufacturer. For all parameters except monocytes (r=0.65), the analyser results correlated well with reference methods. Compared with the manual count of aggregated reticulocytes, the instrument showed good agreement with a positive bias. The optical platelet count correlated well with the manual chamber count. In conclusion the analyser was found to be highly reliable for the analysis of feline blood samples in a large veterinary laboratory.

Abstract

A novel haematology analyser was evaluated for its use with feline samples. Complete blood cell counts, a five-part differential count, and reticulocyte numbers were determined, and the results compared with reference data. Coefficients of correlation, Passing-Bablok regression analysis and Bland-Altmann difference plots with biases and 95% limits of agreement are reported. Precision and linearity were also studied. The instrument demonstrated very low imprecision, and the tested range of linearity exceeded the reference ranges provided by the manufacturer. For all parameters except monocytes (r=0.65), the analyser results correlated well with reference methods. Compared with the manual count of aggregated reticulocytes, the instrument showed good agreement with a positive bias. The optical platelet count correlated well with the manual chamber count. In conclusion the analyser was found to be highly reliable for the analysis of feline blood samples in a large veterinary laboratory.

Statistics

Citations

27 citations in Web of Science®
28 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

326 downloads since deposited on 24 Mar 2010
20 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Department of Farm Animals
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Date:2011
Deposited On:24 Mar 2010 14:11
Last Modified:05 Apr 2016 14:02
Publisher:Elsevier
ISSN:1090-0233
Publisher DOI:https://doi.org/10.1016/j.tvjl.2010.01.005
PubMed ID:20133174

Download

Preview Icon on Download
Preview
Content: Accepted Version
Filetype: PDF
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations