Objective estimation of visual Acuity with preferential looking


Sturm, V; Cassel, D; Eizenman, M (2011). Objective estimation of visual Acuity with preferential looking. Investigative Ophthalmology and Visual Science, 52(2):708-713.

Abstract

Purpose: A novel Preferential Looking (PL) procedure that uses quantitative analysis of visual scanning parameters is presented.

Methods: Nine adult subjects were presented with a set of 14 visual stimuli (stimuli included 3 uniform grey fields and 1 field with black and white square wave gratings) spanning the range of spatial frequencies from 1.5 cycles/degree to 35.1 cycles/degree (1.3 logMAR to -0.07 logMAR). A remote gaze-tracking system was used to monitor the subject's eye movements and the relative fixation time (RFT) on the grating target. Subsequently, a four alternative forced-choice psychophysical test (4AFC) was performed with the same visual stimuli.

Results: For visual stimuli for which the gratings' positions in the 4AFC test were identified correctly in 100% of the trials ("reliably discriminated"), the mean RFT was 72.5% ± 9.0%. For stimuli for which the spatial frequencies were higher than the subject's psychophysically determined VA threshold ("non-discriminated"), the mean RFT was 25.3% ± 8.5%. Using three repeated trials at each spatial frequency and a VA detector that is based on the conditional probability density functions of the RFT, the average VA was underestimated by 0.06 logMAR (range: 0.00 logMAR to 0.20 logMAR).

Conclusions: In adults, automated quantitative analysis of visual scanning patterns can be used to estimate objectively and rapidly (210 seconds) VA with a mean error of 0.06 logMAR. The novel approach may form the basis for PL procedures that are more objective and accurate than the traditional clinical PL procedures.

Abstract

Purpose: A novel Preferential Looking (PL) procedure that uses quantitative analysis of visual scanning parameters is presented.

Methods: Nine adult subjects were presented with a set of 14 visual stimuli (stimuli included 3 uniform grey fields and 1 field with black and white square wave gratings) spanning the range of spatial frequencies from 1.5 cycles/degree to 35.1 cycles/degree (1.3 logMAR to -0.07 logMAR). A remote gaze-tracking system was used to monitor the subject's eye movements and the relative fixation time (RFT) on the grating target. Subsequently, a four alternative forced-choice psychophysical test (4AFC) was performed with the same visual stimuli.

Results: For visual stimuli for which the gratings' positions in the 4AFC test were identified correctly in 100% of the trials ("reliably discriminated"), the mean RFT was 72.5% ± 9.0%. For stimuli for which the spatial frequencies were higher than the subject's psychophysically determined VA threshold ("non-discriminated"), the mean RFT was 25.3% ± 8.5%. Using three repeated trials at each spatial frequency and a VA detector that is based on the conditional probability density functions of the RFT, the average VA was underestimated by 0.06 logMAR (range: 0.00 logMAR to 0.20 logMAR).

Conclusions: In adults, automated quantitative analysis of visual scanning patterns can be used to estimate objectively and rapidly (210 seconds) VA with a mean error of 0.06 logMAR. The novel approach may form the basis for PL procedures that are more objective and accurate than the traditional clinical PL procedures.

Statistics

Citations

Dimensions.ai Metrics
10 citations in Web of Science®
11 citations in Scopus®
12 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

76 downloads since deposited on 24 Dec 2010
10 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Ophthalmology Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2011
Deposited On:24 Dec 2010 09:08
Last Modified:17 Feb 2018 13:05
Publisher:Association for Research in Vision and Ophthalmology
ISSN:0146-0404
OA Status:Hybrid
Publisher DOI:https://doi.org/10.1167/iovs.09-4911
PubMed ID:20861485

Download

Download PDF  'Objective estimation of visual Acuity with preferential looking'.
Preview
Content: Accepted Version
Filetype: PDF
Size: 1MB
View at publisher