Failure modes of Y-TZP crowns at different cusp inclines


Bonfante, E A; Sailer, I; Silva, N R F A; Thompson, V P; Dianne Rekow, E; Coelho, P G (2010). Failure modes of Y-TZP crowns at different cusp inclines. Journal of Dentistry, 38(9):707-712.

Abstract

OBJECTIVES: To compare the reliability of the disto-facial (DF) and mesio-lingual (ML) cusps of an anatomically correct zirconia (Y-TZP) crown system. The research hypotheses tested were: (1) fatigue reliability and failure mode are similar for the ML and DF cusps; (2) failure mode of one cusp does not affect the failure of the other.

METHODS: The average dimensions of a mandibular first molar crown were imported into CAD software; a tooth preparation was modelled by 1.5 mm marginal high reduction of proximal walls and occlusal surface by 2.0 mm. The CAD-based tooth preparation was milled and used as a die to fabricate crowns (n=14) with porcelain veneer on a 0.5 mm Y-TZP core. Crowns were cemented on composite reproductions of the tooth preparation. The crowns were step-stress mouth motion fatigued with sliding (0.7 mm) a tungsten-carbide indenter of 6.25 mm diameter down on the inclines of either the DF or ML cusps. Use level probability Weibull curve with use stress of 200 N and the reliability for completion of a mission of 50,000 cycles at 200 N load were calculated.

RESULTS: Reliability for a 200 N at 50,000 cycles mission was not different between tested cusps. SEM imaging showed large cohesive failures within the veneer for the ML and smaller for the DF. Fractures originated from the contact area regardless of the cusp loaded.

CONCLUSION: No significant difference on fatigue reliability was observed between the DF compared to the ML cusp. Fracture of one cusp did not affect the other.

Abstract

OBJECTIVES: To compare the reliability of the disto-facial (DF) and mesio-lingual (ML) cusps of an anatomically correct zirconia (Y-TZP) crown system. The research hypotheses tested were: (1) fatigue reliability and failure mode are similar for the ML and DF cusps; (2) failure mode of one cusp does not affect the failure of the other.

METHODS: The average dimensions of a mandibular first molar crown were imported into CAD software; a tooth preparation was modelled by 1.5 mm marginal high reduction of proximal walls and occlusal surface by 2.0 mm. The CAD-based tooth preparation was milled and used as a die to fabricate crowns (n=14) with porcelain veneer on a 0.5 mm Y-TZP core. Crowns were cemented on composite reproductions of the tooth preparation. The crowns were step-stress mouth motion fatigued with sliding (0.7 mm) a tungsten-carbide indenter of 6.25 mm diameter down on the inclines of either the DF or ML cusps. Use level probability Weibull curve with use stress of 200 N and the reliability for completion of a mission of 50,000 cycles at 200 N load were calculated.

RESULTS: Reliability for a 200 N at 50,000 cycles mission was not different between tested cusps. SEM imaging showed large cohesive failures within the veneer for the ML and smaller for the DF. Fractures originated from the contact area regardless of the cusp loaded.

CONCLUSION: No significant difference on fatigue reliability was observed between the DF compared to the ML cusp. Fracture of one cusp did not affect the other.

Statistics

Citations

Dimensions.ai Metrics
29 citations in Web of Science®
26 citations in Scopus®
25 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 15 Feb 2011
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Fixed and Removable Prosthodontics
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2010
Deposited On:15 Feb 2011 07:29
Last Modified:17 Feb 2018 18:38
Publisher:Elsevier
ISSN:0300-5712
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.jdent.2010.04.001
PubMed ID:20382197

Download