Thermal spread of vessel-sealing devices evaluated in a clinically relevant in vitro model


Eberli, D; Hefermehl, L J; Müller, A; Sulser, T; Knönagel, H (2011). Thermal spread of vessel-sealing devices evaluated in a clinically relevant in vitro model. Urologia Internationalis, 86(4):476-482.

Abstract

Introduction: Bipolar vessel-sealing devices (VSDs) have advantages in urological surgeries (less hemorrhage, shorter operating time). However, these instruments can cause thermal injuries, which can result in neural damage and necrosis. The objectives of this study were to establish a reproducible in vitro model for standardized assessment of electrosurgical devices and to evaluate whether optimized placement of surgical instruments can reduce the thermal spread. Methods: We evaluated thermal spread of two VSDs in vitro using thin bovine muscle strips. Thermal injury was measured using an infrared camera, temperature probes and histology. The recordings were made with the VSD alone and with a rectangular clamp next to the VSD. Results: Both instruments showed a significant temperature spread of 2.5 mm lateral to the VSD. The placement of a metal clamp next to the VSD significantly reduced the temperature spread. Histological examinations were able to underline these findings. Conclusions: In this study we describe a straightforward clinically relevant in vitro model for the evaluation of future electrosurgical instruments. We demonstrated that the thermal spread of VSD could be further reduced by optimized placement of an additional surgical instrument. Our results could help surgeons protect sensitive structures like nerves in the vicinity of the VSD.

Abstract

Introduction: Bipolar vessel-sealing devices (VSDs) have advantages in urological surgeries (less hemorrhage, shorter operating time). However, these instruments can cause thermal injuries, which can result in neural damage and necrosis. The objectives of this study were to establish a reproducible in vitro model for standardized assessment of electrosurgical devices and to evaluate whether optimized placement of surgical instruments can reduce the thermal spread. Methods: We evaluated thermal spread of two VSDs in vitro using thin bovine muscle strips. Thermal injury was measured using an infrared camera, temperature probes and histology. The recordings were made with the VSD alone and with a rectangular clamp next to the VSD. Results: Both instruments showed a significant temperature spread of 2.5 mm lateral to the VSD. The placement of a metal clamp next to the VSD significantly reduced the temperature spread. Histological examinations were able to underline these findings. Conclusions: In this study we describe a straightforward clinically relevant in vitro model for the evaluation of future electrosurgical instruments. We demonstrated that the thermal spread of VSD could be further reduced by optimized placement of an additional surgical instrument. Our results could help surgeons protect sensitive structures like nerves in the vicinity of the VSD.

Statistics

Citations

10 citations in Web of Science®
10 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

786 downloads since deposited on 04 Aug 2011
87 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Division of Surgical Research
04 Faculty of Medicine > University Hospital Zurich > Urological Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2011
Deposited On:04 Aug 2011 11:59
Last Modified:24 Jun 2016 13:11
Publisher:Karger
ISSN:0042-1138
Additional Information:© 2010 S. Karger AG
Publisher DOI:https://doi.org/10.1159/000324253
PubMed ID:21525721