Header

UZH-Logo

Maintenance Infos

Rivaroxaban: Quantification by anti-FXa assay and influence on coagulation tests A study in 9 Swiss laboratories


Asmis, L M; Alberio, L; Angelillo-Scherrer, A; Korte, W; Mendez, A; Reber, G; Seifert, Burkhardt; Stricker, H; Tsakiris, D A; Wuillemin, W A (2012). Rivaroxaban: Quantification by anti-FXa assay and influence on coagulation tests A study in 9 Swiss laboratories. Thrombosis Research, 129(4):492-498.

Abstract

INTRODUCTION: Rivaroxaban (RXA) is licensed for prophylaxis of venous thromboembolism after major orthopaedic surgery of the lower limbs. Currently, no test to quantify RXA in plasma has been validated in an inter-laboratory setting. Our study had three aims: to assess i) the feasibility of RXA quantification with a commercial anti-FXa assay, ii) its accuracy and precision in an inter-laboratory setting, and iii) the influence of 10mg of RXA on routine coagulation tests. METHODS: The same chromogenic anti-FXa assay (Hyphen BioMed) was used in all participating laboratories. RXA calibrators and sets of blinded probes (aim ii.) were prepared in vitro by spiking normal plasma. The precise RXA content was assessed by high-pressure liquid chromatography-tandem mass spectrometry. For ex-vivo studies (aim iii), plasma samples from 20 healthy volunteers taken before and 2 - 3hours after ingestion of 10mg of RXA were analyzed by participating laboratories. RESULTS: RXA can be assayed chromogenically. Among the participating laboratories, the mean accuracy and the mean coefficient of variation for precision of RXA quantification were 7.0% and 8.8%, respectively. Mean RXA concentration was 114±43μg/L .RXA significantly altered prothrombin time, activated partial thromboplastin time, factor analysis for intrinsic and extrinsic factors. Determinations of thrombin time, fibrinogen, FXIII and D-Dimer levels were not affected. CONCLUSIONS: RXA plasma levels can be quantified accurately and precisely by a chromogenic anti-FXa assay on different coagulometers in different laboratories. Ingestion of 10mg RXA results in significant alterations of both PT- and aPTT-based coagulation assays.

Abstract

INTRODUCTION: Rivaroxaban (RXA) is licensed for prophylaxis of venous thromboembolism after major orthopaedic surgery of the lower limbs. Currently, no test to quantify RXA in plasma has been validated in an inter-laboratory setting. Our study had three aims: to assess i) the feasibility of RXA quantification with a commercial anti-FXa assay, ii) its accuracy and precision in an inter-laboratory setting, and iii) the influence of 10mg of RXA on routine coagulation tests. METHODS: The same chromogenic anti-FXa assay (Hyphen BioMed) was used in all participating laboratories. RXA calibrators and sets of blinded probes (aim ii.) were prepared in vitro by spiking normal plasma. The precise RXA content was assessed by high-pressure liquid chromatography-tandem mass spectrometry. For ex-vivo studies (aim iii), plasma samples from 20 healthy volunteers taken before and 2 - 3hours after ingestion of 10mg of RXA were analyzed by participating laboratories. RESULTS: RXA can be assayed chromogenically. Among the participating laboratories, the mean accuracy and the mean coefficient of variation for precision of RXA quantification were 7.0% and 8.8%, respectively. Mean RXA concentration was 114±43μg/L .RXA significantly altered prothrombin time, activated partial thromboplastin time, factor analysis for intrinsic and extrinsic factors. Determinations of thrombin time, fibrinogen, FXIII and D-Dimer levels were not affected. CONCLUSIONS: RXA plasma levels can be quantified accurately and precisely by a chromogenic anti-FXa assay on different coagulometers in different laboratories. Ingestion of 10mg RXA results in significant alterations of both PT- and aPTT-based coagulation assays.

Statistics

Citations

Dimensions.ai Metrics
94 citations in Web of Science®
112 citations in Scopus®
175 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 11 Jan 2012
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Epidemiology, Biostatistics and Prevention Institute (EBPI)
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2012
Deposited On:11 Jan 2012 12:36
Last Modified:16 Feb 2018 23:10
Publisher:Elsevier
ISSN:0049-3848
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.thromres.2011.06.031
PubMed ID:21840043

Download