Header

UZH-Logo

Maintenance Infos

Comparison of FBAR and QCM-D sensitivity dependence on adlayer thickness and viscosity


Nirschl, M; Schreiter, M; Vörös, J (2011). Comparison of FBAR and QCM-D sensitivity dependence on adlayer thickness and viscosity. Sensors and Actuators A: Physical, 165(2):415-421.

Abstract

Unlike the quartz crystal microbalance, which has been used extensively for the analysis of biochemical interactions, only few measurements with biochemical adsorbent have been done with film bulk acoustic resonators (FBAR). In this paper, the FBAR behaviour on exposure to a lipid vesicle solution and the formation of a polyelectrolyte multilayer structure is investigated and compared with the results obtained with the quartz crystal microbalance. Differences in the resonator response were found between the two techniques and depending on the resonators resonance frequency ranging from the MHz to the GHz regime. As an explanation, we suggest that the penetration depth and the influence on viscoelastic properties, which are both known to be frequency dependent, cause the variations in the results. As a consequence, the higher operating resonance frequencies of the FBAR increase the sensitivity to changes in the viscoelasticity of the adsorbent and also decrease the sensing length of the device.

Abstract

Unlike the quartz crystal microbalance, which has been used extensively for the analysis of biochemical interactions, only few measurements with biochemical adsorbent have been done with film bulk acoustic resonators (FBAR). In this paper, the FBAR behaviour on exposure to a lipid vesicle solution and the formation of a polyelectrolyte multilayer structure is investigated and compared with the results obtained with the quartz crystal microbalance. Differences in the resonator response were found between the two techniques and depending on the resonators resonance frequency ranging from the MHz to the GHz regime. As an explanation, we suggest that the penetration depth and the influence on viscoelastic properties, which are both known to be frequency dependent, cause the variations in the results. As a consequence, the higher operating resonance frequencies of the FBAR increase the sensitivity to changes in the viscoelasticity of the adsorbent and also decrease the sensing length of the device.

Statistics

Citations

15 citations in Web of Science®
15 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

353 downloads since deposited on 12 Jan 2012
54 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Language:English
Date:2011
Deposited On:12 Jan 2012 17:22
Last Modified:05 Apr 2016 15:18
Publisher:Elsevier
ISSN:0924-4247
Publisher DOI:https://doi.org/10.1016/j.sna.2010.11.003

Download

Preview Icon on Download
Preview
Content: Accepted Version
Filetype: PDF
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations