Performance evaluation of the Sysmex pocH-100iV Diff hematology analyzer for analysis of canine, feline, equine, and bovine blood


Riond, B; Weissenbacher, S; Hofmann-Lehmann, R; Lutz, H (2011). Performance evaluation of the Sysmex pocH-100iV Diff hematology analyzer for analysis of canine, feline, equine, and bovine blood. Veterinary Clinical Pathology, 40(4):484-495.

Abstract

BACKGROUND:

The Sysmex pocH-100iV Diff is an impedance hematology analyzer recently introduced for point-of-care use in veterinary practices in Europe.
OBJECTIVE:

The purpose of this study was to validate the pocH-100iV Diff for analysis of blood samples from dogs, cats, horses, and cattle.
METHODS:

Fresh EDTA-blood samples from healthy and ill dogs (115), cats (94), horses (91), and cattle (78) were analyzed on the pocH-100iV Diff and the Cell-Dyn 3500. Results of the automated WBC differential counts were compared with the manual differential counts for 77 dogs, 65 cats, 40 horses, and 46 cattle. HCT were compared with PCVs obtained by microhematocrit centrifugation. Furthermore, precision, linearity, carry-over, cell aging, and clinical relevance of the pocH-100iV Diff results were assessed.
RESULTS:

Most of the CBC results obtained by the pocH-100iV Diff correlated well with those of the Cell-Dyn 3500. Slightly low correlation was observed for canine MCV and hemoglobin concentration. Lymphocytes correlated well in horses and cattle, but less well in cats and dogs. The mixed cell population termed "OTHRS" (all granulocytes and monocytes for horses and cattle; neutrophils, monocytes, and basophils for cats and dogs) correlated well in all tested species. The instrument overestimated feline and canine eosinophils. In cats, platelet counts showed a strong negative bias.
CONCLUSIONS:

The overall performance of the pocH-100iV Diff was excellent with the noted limitations. The automated differential count can be used as screening tool in conjunction with evaluation of a blood smear.

Abstract

BACKGROUND:

The Sysmex pocH-100iV Diff is an impedance hematology analyzer recently introduced for point-of-care use in veterinary practices in Europe.
OBJECTIVE:

The purpose of this study was to validate the pocH-100iV Diff for analysis of blood samples from dogs, cats, horses, and cattle.
METHODS:

Fresh EDTA-blood samples from healthy and ill dogs (115), cats (94), horses (91), and cattle (78) were analyzed on the pocH-100iV Diff and the Cell-Dyn 3500. Results of the automated WBC differential counts were compared with the manual differential counts for 77 dogs, 65 cats, 40 horses, and 46 cattle. HCT were compared with PCVs obtained by microhematocrit centrifugation. Furthermore, precision, linearity, carry-over, cell aging, and clinical relevance of the pocH-100iV Diff results were assessed.
RESULTS:

Most of the CBC results obtained by the pocH-100iV Diff correlated well with those of the Cell-Dyn 3500. Slightly low correlation was observed for canine MCV and hemoglobin concentration. Lymphocytes correlated well in horses and cattle, but less well in cats and dogs. The mixed cell population termed "OTHRS" (all granulocytes and monocytes for horses and cattle; neutrophils, monocytes, and basophils for cats and dogs) correlated well in all tested species. The instrument overestimated feline and canine eosinophils. In cats, platelet counts showed a strong negative bias.
CONCLUSIONS:

The overall performance of the pocH-100iV Diff was excellent with the noted limitations. The automated differential count can be used as screening tool in conjunction with evaluation of a blood smear.

Statistics

Citations

8 citations in Web of Science®
7 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Department of Farm Animals
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Date:2011
Deposited On:01 Feb 2012 09:46
Last Modified:05 Apr 2016 15:21
Publisher:Wiley-Blackwell
ISSN:0275-6382
Publisher DOI:https://doi.org/10.1111/j.1939-165X.2011.00372.x
PubMed ID:22092369

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations