Abstract

Objectives: The aim of the present multi-center study was to evaluate the treatment outcome of immediately restored one-piece single-tooth implants with a diameter of 3 mm after 1 year.

Material and methods: A total of 57 one-piece implants (NobelDirect® 3.0) were inserted in 47 patients (26 females, 21 males) with a mean age of 31 years (range: 17–76 years) at five different centers. The implants replaced maxillary lateral incisors and mandibular incisors. The implants were placed either in conjunction with tooth extraction or in healed sites, and all implants were immediately restored with a provisional resin crown. If needed, the abutment part of the implant was prepared before crown cementation. The permanent crown was placed after 1.9–14.5 months. Radiographs were taken at implant insertion as well as after 6 and 12 months to evaluate the peri-implant marginal bone level and bone loss. Moreover, plaque, bleeding on probing and complications were assessed.

Results: A total of 44 patients (23 females, 21 males) with 54 implants were available for the 1-year follow-up. One implant was lost, thus the 1-year implant survival was 98%. A statistically significant mean marginal bone loss was observed between baseline and 6 months (1.1 mm, range: −0.7 to 4.4 mm; n=49) and between baseline and 12 months (1.6 mm, range: −0.8 to 4.6 mm; n=50). A total of 18% of the implants were characterized by a bone loss of more than 3 mm. No bleeding on probing was observed around 83% of the implants. Plaque was registered at 15% of the implants. The most common complications were related to the provisional crown, i.e. fracture (n=3) and loss of retention (n=3).

Conclusions: A high 1-year implant survival was observed in the present study. However, the excessive peri-implant marginal bone loss around several implants indicates that this implant should be used with caution until further studies have been conducted.

To cite this article: Zembić A, Johannesen LH, Schou S, Malo P, Reichert T, Farella M, Hämmerle CHF. Immediately restored one-piece single-tooth implants with reduced diameter: one-year results of a multi-center stud

Abstract

Objectives: The aim of the present multi-center study was to evaluate the treatment outcome of immediately restored one-piece single-tooth implants with a diameter of 3 mm after 1 year.

Material and methods: A total of 57 one-piece implants (NobelDirect® 3.0) were inserted in 47 patients (26 females, 21 males) with a mean age of 31 years (range: 17–76 years) at five different centers. The implants replaced maxillary lateral incisors and mandibular incisors. The implants were placed either in conjunction with tooth extraction or in healed sites, and all implants were immediately restored with a provisional resin crown. If needed, the abutment part of the implant was prepared before crown cementation. The permanent crown was placed after 1.9–14.5 months. Radiographs were taken at implant insertion as well as after 6 and 12 months to evaluate the peri-implant marginal bone level and bone loss. Moreover, plaque, bleeding on probing and complications were assessed.

Results: A total of 44 patients (23 females, 21 males) with 54 implants were available for the 1-year follow-up. One implant was lost, thus the 1-year implant survival was 98%. A statistically significant mean marginal bone loss was observed between baseline and 6 months (1.1 mm, range: −0.7 to 4.4 mm; n=49) and between baseline and 12 months (1.6 mm, range: −0.8 to 4.6 mm; n=50). A total of 18% of the implants were characterized by a bone loss of more than 3 mm. No bleeding on probing was observed around 83% of the implants. Plaque was registered at 15% of the implants. The most common complications were related to the provisional crown, i.e. fracture (n=3) and loss of retention (n=3).

Conclusions: A high 1-year implant survival was observed in the present study. However, the excessive peri-implant marginal bone loss around several implants indicates that this implant should be used with caution until further studies have been conducted.

To cite this article: Zembić A, Johannesen LH, Schou S, Malo P, Reichert T, Farella M, Hämmerle CHF. Immediately restored one-piece single-tooth implants with reduced diameter: one-year results of a multi-center stud

Statistics

Citations

21 citations in Web of Science®
19 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

1 download since deposited on 22 Feb 2012
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Fixed and Removable Prosthodontics
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:23 January 2012
Deposited On:22 Feb 2012 16:57
Last Modified:07 Dec 2017 12:14
Publisher:Wiley-Blackwell
ISSN:0905-7161
Publisher DOI:https://doi.org/10.1111/j.1600-0501.2011.02174.x
PubMed ID:21504480

Download