Commissioning of photon beams of a flattening filter-free linear accelerator and the accuracy of beam modeling using an anisotropic analytical algorithm


Hrbacek, J; Lang, S; Klöck, S (2011). Commissioning of photon beams of a flattening filter-free linear accelerator and the accuracy of beam modeling using an anisotropic analytical algorithm. International Journal of Radiation Oncology, Biology, Physics, 80(4):1228-1237.

Abstract

PURPOSE:

To investigate dosimetric characteristics of a new linear accelerator designed to deliver flattened, as well as flattening filter-free (FFF), beams. To evaluate the accuracy of beam modeling under physical conditions using an anisotropic analytical algorithm.
METHODS AND MATERIALS:

Dosimetric data including depth dose curves, profiles, surface dose, penumbra, out-of-field dose, output, total and scatter factors were examined for four beams (X6, X6FFF, X10, and X10FFF) of Varian's TrueBeam machine. Beams modeled by anisotropic analytical algorithm were compared with measured dataset.
RESULTS:

FFF beams have lower mean energy (tissue-phantom ratio at the depths of 20 and 10 cm (TPR 20/10): X6, 0.667; X6FFF, 0.631; X10, 0.738; X10FFF, 0.692); maximum dose is located closer to the surface; and surface dose increases by 10%. FFF profiles have sharper but faster diverging penumbra. For small fields and shallow depths, dose outside a field is lower for FFF beams; however, the advantage fades with increasing phantom scatter. Output increases 2.26 times for X6FFF and 4.03 times for X10FFF and is less variable with field size; collimator exchange effect is reduced. A good agreement between modeled and measured data is observed. Criteria of 2% depth-dose and 2-mm distance-to-agreement are always met.
CONCLUSION:

Reference dosimetric characteristics of TrueBeam photon bundles were obtained, and successful modeling of the beams was achieved.

Copyright © 2011 Elsevier Inc. All rights reserved.

Abstract

PURPOSE:

To investigate dosimetric characteristics of a new linear accelerator designed to deliver flattened, as well as flattening filter-free (FFF), beams. To evaluate the accuracy of beam modeling under physical conditions using an anisotropic analytical algorithm.
METHODS AND MATERIALS:

Dosimetric data including depth dose curves, profiles, surface dose, penumbra, out-of-field dose, output, total and scatter factors were examined for four beams (X6, X6FFF, X10, and X10FFF) of Varian's TrueBeam machine. Beams modeled by anisotropic analytical algorithm were compared with measured dataset.
RESULTS:

FFF beams have lower mean energy (tissue-phantom ratio at the depths of 20 and 10 cm (TPR 20/10): X6, 0.667; X6FFF, 0.631; X10, 0.738; X10FFF, 0.692); maximum dose is located closer to the surface; and surface dose increases by 10%. FFF profiles have sharper but faster diverging penumbra. For small fields and shallow depths, dose outside a field is lower for FFF beams; however, the advantage fades with increasing phantom scatter. Output increases 2.26 times for X6FFF and 4.03 times for X10FFF and is less variable with field size; collimator exchange effect is reduced. A good agreement between modeled and measured data is observed. Criteria of 2% depth-dose and 2-mm distance-to-agreement are always met.
CONCLUSION:

Reference dosimetric characteristics of TrueBeam photon bundles were obtained, and successful modeling of the beams was achieved.

Copyright © 2011 Elsevier Inc. All rights reserved.

Statistics

Citations

72 citations in Web of Science®
75 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 28 Feb 2012
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Radiation Oncology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2011
Deposited On:28 Feb 2012 16:19
Last Modified:05 Apr 2016 15:42
Publisher:Elsevier
ISSN:1879-355X
Publisher DOI:https://doi.org/10.1016/j.ijrobp.2010.09.050
PubMed ID:21129855

Download

Preview Icon on Download
Content: Published Version
Filetype: PDF - Registered users only
Size: 1MB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations