Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements


Buck, Florian M; Guggenberger, Roman; Koch, Peter P; Pfirrmann, Christian W A (2012). Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements. American Journal of Roentgenology, 199(5):W607-W612.

Abstract

OBJECTIVE: The purpose of this study was to evaluate the interchangeability of femoral and tibial torsion measurements obtained with 3D models based on low-dose biplanar radiographs and standard CT measurements by testing the following two hypotheses: that there is excellent agreement between the two methods and that there is excellent interreader agreement.
MATERIALS AND METHODS: Two independent readers used 3D models based on low-dose simultaneous biplanar radiographs and axial CT images to measure femoral and tibial torsion in 35 patients (mean age, 65 years; range, 46-89 years) with osteoarthritis of the knee who were to undergo prosthesis insertion. The two measurements were compared by means of Bland-Altman plots and descriptive statistics. Interreader agreement was quantified with intraclass correlation coefficients.
RESULTS: The average differences between readers on the CT measurements were 1.3° (range, 0°-11°) for the femur and 1.5° (range, 0°-12°) for the tibia. The average differences for the measurements obtained with the 3D model were 0.1° (range, 0°-9°) for the femur and 0.8° (range, 0°-10°) for the tibia. The average differences between the two methods were 0° (range, -5° to 7°) for the femoral measurements and 3° (range, -12° to 5°) for the tibial measurements. Bland-Altman plots showed no relevant differences between the results of the two measurement modalities. Except for one measurement of femoral torsion and one measurement of tibial torsion, all results based on the 3D models were within the 95% limit of agreement (mean ± 1.96 SD). Interreader agreement was statistically significant (p < 0.001) for all measurements with high intraclass correlation coefficients (> 0.9).
CONCLUSION: Femoral and tibial torsion measurements obtained with 3D models based on biplanar radiographs are interchangeable with standard CT measurements in patients with osteoarthritis of the knee.

Abstract

OBJECTIVE: The purpose of this study was to evaluate the interchangeability of femoral and tibial torsion measurements obtained with 3D models based on low-dose biplanar radiographs and standard CT measurements by testing the following two hypotheses: that there is excellent agreement between the two methods and that there is excellent interreader agreement.
MATERIALS AND METHODS: Two independent readers used 3D models based on low-dose simultaneous biplanar radiographs and axial CT images to measure femoral and tibial torsion in 35 patients (mean age, 65 years; range, 46-89 years) with osteoarthritis of the knee who were to undergo prosthesis insertion. The two measurements were compared by means of Bland-Altman plots and descriptive statistics. Interreader agreement was quantified with intraclass correlation coefficients.
RESULTS: The average differences between readers on the CT measurements were 1.3° (range, 0°-11°) for the femur and 1.5° (range, 0°-12°) for the tibia. The average differences for the measurements obtained with the 3D model were 0.1° (range, 0°-9°) for the femur and 0.8° (range, 0°-10°) for the tibia. The average differences between the two methods were 0° (range, -5° to 7°) for the femoral measurements and 3° (range, -12° to 5°) for the tibial measurements. Bland-Altman plots showed no relevant differences between the results of the two measurement modalities. Except for one measurement of femoral torsion and one measurement of tibial torsion, all results based on the 3D models were within the 95% limit of agreement (mean ± 1.96 SD). Interreader agreement was statistically significant (p < 0.001) for all measurements with high intraclass correlation coefficients (> 0.9).
CONCLUSION: Femoral and tibial torsion measurements obtained with 3D models based on biplanar radiographs are interchangeable with standard CT measurements in patients with osteoarthritis of the knee.

Statistics

Citations

22 citations in Web of Science®
26 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2012
Deposited On:30 Nov 2012 11:27
Last Modified:05 Apr 2016 16:06
Publisher:American Roentgen Ray Society
ISSN:0361-803X
Publisher DOI:https://doi.org/10.2214/AJR.11.8295
PubMed ID:23096205

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations