Feasibility of the Engager™ aortic transcatheter valve system using a flexible over-the-wire design


Sündermann, Simon H; Grünenfelder, Jürg; Corti, Roberto; Rastan, Ardawan J; Linke, Axel; Lange, Rüdiger; Falk, Volkmar; Bleiziffer, Sabine (2012). Feasibility of the Engager™ aortic transcatheter valve system using a flexible over-the-wire design. European Journal of Cardio-Thoracic Surgery, 42(4):e48-e52.

Abstract

OBJECTIVES: The aim was to investigate the safety and feasibility of the redesigned Engager™ transcatheter aortic valve implantation (TAVI) system.
METHODS: Transapical aortic valve implantation with the Engager™ valve prosthesis was intended in 11 patients, and performed in 10. Endpoints were defined according to the valve academic research consortium recommendations for reporting outcomes of TAVI in clinical trials.
RESULTS: All 10 patients were implanted successfully. No devicerelated or delivery system complications like coronary obstruction or aortic dissection emerged. One patient (10%) died from non-device-related reasons at post-operative day 23 of multi-organ failure. The invasively measured peak-to-peak gradient after valve implantation was 7.1 ± 3.5 mmHg. In 90%, there was no or only trivial (≤grad I) aortic regurgitation due to paravalvular leakage. In 10% of the patients, aortic regurgitation grade I-II was observed. At 30-day follow up, the mean gradient was 15.6 ± 4.9 mmHg, and no more than a mild transvalvular and paravalvular aortic regurgitation was seen as assessed by transthoracic echocardiography.
CONCLUSIONS: Application of the Engager™ TAVI system is safe and feasible. Prosthesis deployment in an anatomically correct position was facilitated by the design of the valve prosthesis and was successful in all patients. No device or delivery-system-related complications emerged. Safety and feasibility endpoints were met. Good results concerning the aortic valve performance after implantation and at 30-day follow up were ascertained. These results encouraged the start of a European Pivotal trial including patients to date.

Abstract

OBJECTIVES: The aim was to investigate the safety and feasibility of the redesigned Engager™ transcatheter aortic valve implantation (TAVI) system.
METHODS: Transapical aortic valve implantation with the Engager™ valve prosthesis was intended in 11 patients, and performed in 10. Endpoints were defined according to the valve academic research consortium recommendations for reporting outcomes of TAVI in clinical trials.
RESULTS: All 10 patients were implanted successfully. No devicerelated or delivery system complications like coronary obstruction or aortic dissection emerged. One patient (10%) died from non-device-related reasons at post-operative day 23 of multi-organ failure. The invasively measured peak-to-peak gradient after valve implantation was 7.1 ± 3.5 mmHg. In 90%, there was no or only trivial (≤grad I) aortic regurgitation due to paravalvular leakage. In 10% of the patients, aortic regurgitation grade I-II was observed. At 30-day follow up, the mean gradient was 15.6 ± 4.9 mmHg, and no more than a mild transvalvular and paravalvular aortic regurgitation was seen as assessed by transthoracic echocardiography.
CONCLUSIONS: Application of the Engager™ TAVI system is safe and feasible. Prosthesis deployment in an anatomically correct position was facilitated by the design of the valve prosthesis and was successful in all patients. No device or delivery-system-related complications emerged. Safety and feasibility endpoints were met. Good results concerning the aortic valve performance after implantation and at 30-day follow up were ascertained. These results encouraged the start of a European Pivotal trial including patients to date.

Statistics

Citations

14 citations in Web of Science®
18 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

1 download since deposited on 15 Feb 2013
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiovascular Surgery
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2012
Deposited On:15 Feb 2013 09:48
Last Modified:05 Apr 2016 16:27
Publisher:Elsevier
ISSN:1010-7940
Publisher DOI:https://doi.org/10.1093/ejcts/ezs389
PubMed ID:22743081

Download