Is the PASCAL-Tonometer suitable for measuring intraocular pressure in clinical routine? Long- and short-term reproducibility of dynamic contour tonometry


Herdener, S; Hafizovic, D; Pache, M; Lautebach, S; Funk, J (2008). Is the PASCAL-Tonometer suitable for measuring intraocular pressure in clinical routine? Long- and short-term reproducibility of dynamic contour tonometry. European Journal of Ophthalmology, 18(1):39-43.

Abstract

PURPOSE: Dynamic contour tonometry (DCT) is a new technique to measure intraocular pressure (IOP). In several studies no correlation between IOP values and corneal thickness was shown with DCT. This is in contrast to the gold standard, Goldmann applanation tonometry (GAT). The authors evaluated the reproducibility (RP) of DCT compared to GAT, a prerequisite for its introduction into clinical routine. METHODS: IOP was measured with both DCT and GAT in 50 subjects with normal cornea. To evaluate the short-term RP, four DCT and four GAT measurements were performed at day 1 in a randomized order. Long-term RP was determined by one additional measurement per method at day 2, 5, and 8. RESULTS: The short-term RP was defined as the mean value of all standard deviations (SD) of the individual measurements at day 1. Short-term RP was 1.1 mmHg for GAT and 1.2 mmHg for DCT. For long-term reproducibility, mean SD was 1.2 mmHg for GAT and 1.5 mmHg for DCT. Bland-Altman revealed a good agreement of the two methods. However, mean DCT values were on average 0.8+/-1.1 mmHg higher than GAT values. A significant correlation was observed between GAT and CCT (r2=0.15, p=0.006), but not between DCT and CCT (r2=0.032, p=0.21). CONCLUSIONS: Short- and long-term reproducibility of DCT is comparable to that of GAT. GAT is more affected by CCT than DCT, measuring higher IOPs in eyes with higher central corneal thickness.

Abstract

PURPOSE: Dynamic contour tonometry (DCT) is a new technique to measure intraocular pressure (IOP). In several studies no correlation between IOP values and corneal thickness was shown with DCT. This is in contrast to the gold standard, Goldmann applanation tonometry (GAT). The authors evaluated the reproducibility (RP) of DCT compared to GAT, a prerequisite for its introduction into clinical routine. METHODS: IOP was measured with both DCT and GAT in 50 subjects with normal cornea. To evaluate the short-term RP, four DCT and four GAT measurements were performed at day 1 in a randomized order. Long-term RP was determined by one additional measurement per method at day 2, 5, and 8. RESULTS: The short-term RP was defined as the mean value of all standard deviations (SD) of the individual measurements at day 1. Short-term RP was 1.1 mmHg for GAT and 1.2 mmHg for DCT. For long-term reproducibility, mean SD was 1.2 mmHg for GAT and 1.5 mmHg for DCT. Bland-Altman revealed a good agreement of the two methods. However, mean DCT values were on average 0.8+/-1.1 mmHg higher than GAT values. A significant correlation was observed between GAT and CCT (r2=0.15, p=0.006), but not between DCT and CCT (r2=0.032, p=0.21). CONCLUSIONS: Short- and long-term reproducibility of DCT is comparable to that of GAT. GAT is more affected by CCT than DCT, measuring higher IOPs in eyes with higher central corneal thickness.

Statistics

Citations

14 citations in Web of Science®
18 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

3 downloads since deposited on 11 Dec 2008
0 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Ophthalmology Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2008
Deposited On:11 Dec 2008 08:23
Last Modified:05 Apr 2016 12:39
Publisher:Wichtig Editore
ISSN:1120-6721
Official URL:http://www.artificial-organs.com/public/EJO/Article/Articleabstract.aspx?UidArticle=D135A2E1-6139-4D33-BA7F-8FB712AB6C91&t=EJO
PubMed ID:18203083

Download

Preview Icon on Download
Filetype: PDF - Registered users only
Size: 5MB

Article Networks

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations