Networked multielectrode left ventricular pacing lead for avoidance of phrenic nerve stimulation in a canine model


Wecke, Liliane; van Hunnik, Arne; Thompson, Todd; DiCarlo, Lorenzo; Zdeblick, Mark; Auricchio, Angelo; Prinzen, Frits W (2012). Networked multielectrode left ventricular pacing lead for avoidance of phrenic nerve stimulation in a canine model. Heart Rhythm, 9(5):789-795.

Abstract

BACKGROUND: In cardiac resynchronization therapy, left ventricular stimulation may lead to concomitant phrenic nerve stimulation (PNS).
OBJECTIVE: To evaluate a new networked multielectrode lead with 16 electrode segments (SEGs) configured into groups of 4, forming a virtual band (VBAND) around the lead. Each electrode is individually programmable using an embedded integrated circuit.
METHODS: In 8 anesthetized dogs, the lead was positioned in a left ventricular coronary vein. The voltage thresholds for cardiac stimulation and PNS were measured for different electrode configurations, including "VBAND-VBAND" (∼conventional bipolar pacing), "SEG-VBAND", and "SEG-SEG" (anode and cathode within the same VBAND). The measurements were performed (1) with closed chest and (2) after opening the chest and repositioning the phrenic nerve to above the lead, simulating a worst-case scenario.
RESULTS: Compared with the conventional VBAND-VBAND stimulation, the SEG-SEG stimulation increased the PNS threshold and raised the difference between phrenic and cardiac thresholds from 6.2 ± 2.3 to 9.5 ±0.3 V in the closed chest condition and from 1.4 ± 1.6 to 9.0 ± 1.0 V in the worst-case scenario (both P < .001). Both SEG-VBAND and SEG-SEG stimulations reduced the cardiac threshold and increased pacing impedance, thus reducing the required cardiac pacing power by 77%-80% (P <.001 and P <.01 for closed and open chest, respectively).
CONCLUSION: This novel multielectrode pacing lead achieves low cardiac and high extracardiac stimulation thresholds during left ventricular pacing in a canine model. The virtual elimination of PNS may facilitate and improve the application of cardiac resynchronization therapy.

Abstract

BACKGROUND: In cardiac resynchronization therapy, left ventricular stimulation may lead to concomitant phrenic nerve stimulation (PNS).
OBJECTIVE: To evaluate a new networked multielectrode lead with 16 electrode segments (SEGs) configured into groups of 4, forming a virtual band (VBAND) around the lead. Each electrode is individually programmable using an embedded integrated circuit.
METHODS: In 8 anesthetized dogs, the lead was positioned in a left ventricular coronary vein. The voltage thresholds for cardiac stimulation and PNS were measured for different electrode configurations, including "VBAND-VBAND" (∼conventional bipolar pacing), "SEG-VBAND", and "SEG-SEG" (anode and cathode within the same VBAND). The measurements were performed (1) with closed chest and (2) after opening the chest and repositioning the phrenic nerve to above the lead, simulating a worst-case scenario.
RESULTS: Compared with the conventional VBAND-VBAND stimulation, the SEG-SEG stimulation increased the PNS threshold and raised the difference between phrenic and cardiac thresholds from 6.2 ± 2.3 to 9.5 ±0.3 V in the closed chest condition and from 1.4 ± 1.6 to 9.0 ± 1.0 V in the worst-case scenario (both P < .001). Both SEG-VBAND and SEG-SEG stimulations reduced the cardiac threshold and increased pacing impedance, thus reducing the required cardiac pacing power by 77%-80% (P <.001 and P <.01 for closed and open chest, respectively).
CONCLUSION: This novel multielectrode pacing lead achieves low cardiac and high extracardiac stimulation thresholds during left ventricular pacing in a canine model. The virtual elimination of PNS may facilitate and improve the application of cardiac resynchronization therapy.

Statistics

Citations

5 citations in Web of Science®
4 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Cardiocentro Ticino
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2012
Deposited On:28 Feb 2013 16:39
Last Modified:05 Apr 2016 16:37
Publisher:Elsevier
ISSN:1547-5271
Publisher DOI:https://doi.org/10.1016/j.hrthm.2011.11.052
PubMed ID:22138135

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations