First experience with monochromatic coronary computed tomography angiography from a 64-slice CT scanner with Gemstone Spectral Imaging (GSI)


Fuchs, Tobias A; Stehli, Julia; Kaufmann, Philipp A; et al (2013). First experience with monochromatic coronary computed tomography angiography from a 64-slice CT scanner with Gemstone Spectral Imaging (GSI). Journal of Cardiovascular Computed Tomography, 7(1):25-31.

Abstract

BACKGROUND: New technology combining dual-energy CT with the latest gemstone detectors for spectral imaging (GSI) can be used to synthesize monochromatic images that mimic images as if different monochromatic x-ray sources were used. OBJECTIVE: The aim of the study was to evaluate the optimal combination of monochromatic image energy and adaptive statistical iterative reconstruction (ASiR) for monochromatic reconstruction of coronary CT angiography (CTA) images with the use of GSI. METHODS: Twenty consecutive patients underwent coronary CTA on a GSI capable 64-slice CT scanner (Discovery CT 750 High Definition, GE Healthcare). In 7 sets of monochromatic images (60, 65, 70, 75, 80, 90, and 110 keV; each with increasing contributions of ASiR, ie, 0%, 20%, 40%, 60%, and 80%; n = 35 reconstructions per patient), signal-to-noise (aortic root) and contrast-to-noise (left main artery) ratios were assessed. Signal-to-noise ratio, contrast-to-noise ratio, and image quality (graded on a 5-point Likert scale) were assessed in all above monochromatic reconstructions and compared with the respective standard (conventional polychromatic) image. RESULTS: Compared with conventional polychromatic images, reconstructions with 60 keV and 80% ASiR showed the highest improvement in contrast-to-noise (144%; P < 0.001) and signal-to-noise ratio (173%; P < 0.001). Image quality reached a plateau at 65-75 keV with 40%-60% ASiR blending, yielding a maximal image quality score improvement of 50% compared with conventional imaging (P < 0.001). CONCLUSION: In coronary CTA with low radiation technique (mean radiation dose, 1.8 ± 0.7 mSv), GSI with monochromatic reconstructions (65-75 keV) and ASiR (40%-60%) offers significant noise reduction and image quality improvement.

Abstract

BACKGROUND: New technology combining dual-energy CT with the latest gemstone detectors for spectral imaging (GSI) can be used to synthesize monochromatic images that mimic images as if different monochromatic x-ray sources were used. OBJECTIVE: The aim of the study was to evaluate the optimal combination of monochromatic image energy and adaptive statistical iterative reconstruction (ASiR) for monochromatic reconstruction of coronary CT angiography (CTA) images with the use of GSI. METHODS: Twenty consecutive patients underwent coronary CTA on a GSI capable 64-slice CT scanner (Discovery CT 750 High Definition, GE Healthcare). In 7 sets of monochromatic images (60, 65, 70, 75, 80, 90, and 110 keV; each with increasing contributions of ASiR, ie, 0%, 20%, 40%, 60%, and 80%; n = 35 reconstructions per patient), signal-to-noise (aortic root) and contrast-to-noise (left main artery) ratios were assessed. Signal-to-noise ratio, contrast-to-noise ratio, and image quality (graded on a 5-point Likert scale) were assessed in all above monochromatic reconstructions and compared with the respective standard (conventional polychromatic) image. RESULTS: Compared with conventional polychromatic images, reconstructions with 60 keV and 80% ASiR showed the highest improvement in contrast-to-noise (144%; P < 0.001) and signal-to-noise ratio (173%; P < 0.001). Image quality reached a plateau at 65-75 keV with 40%-60% ASiR blending, yielding a maximal image quality score improvement of 50% compared with conventional imaging (P < 0.001). CONCLUSION: In coronary CTA with low radiation technique (mean radiation dose, 1.8 ± 0.7 mSv), GSI with monochromatic reconstructions (65-75 keV) and ASiR (40%-60%) offers significant noise reduction and image quality improvement.

Statistics

Citations

33 citations in Web of Science®
35 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:13 Jun 2013 12:41
Last Modified:07 Dec 2017 21:22
Publisher:Elsevier
ISSN:1876-861X
Publisher DOI:https://doi.org/10.1016/j.jcct.2013.01.004
PubMed ID:23452997

Download

Full text not available from this repository.
View at publisher