Vertebral body stenting versus kyphoplasty for the treatment of osteoporotic vertebral compression fractures: a randomized trial


Werner, Clément M L; Osterhoff, Georg; Schlickeiser, Jannis; Jenni, Raphael; Wanner, Guido A; Ossendorf, Christian; Simmen, Hans-Peter (2013). Vertebral body stenting versus kyphoplasty for the treatment of osteoporotic vertebral compression fractures: a randomized trial. Journal of Bone and Joint Surgery. American Volume, 95(7):577-584.

Abstract

BACKGROUND: In the treatment of vertebral compression fractures, vertebral body stenting with an expandable scaffold inserted before application of the bone cement was developed to impede secondary loss of vertebral height encountered in patients treated with balloon kyphoplasty. The purpose of this study was to clarify whether there are relevant differences between balloon kyphoplasty and vertebral body stenting with regard to perioperative and postoperative findings. METHODS: In a two-armed randomized controlled trial, patients with a total of 100 fresh osteoporotic vertebral compression fractures were treated with either balloon kyphoplasty or vertebral body stenting. The primary outcome was the post-interventional change in the kyphotic angle on radiographs. The secondary outcomes were the maximum pressure of the balloon tamp during inflation, radiation exposure time, perioperative complications, and cement leakage. RESULTS: The mean reduction (and standard deviation) of kyphosis (the kyphotic correction angle) was 4.5° ± 3.6° after balloon kyphoplasty and 4.7° ± 4.2° after vertebral body stenting (p = 0.972). The mean pressures were 24 ± 5 bar (348 ± 72 pounds per square inch [psi]) during vertebral body stenting and 16 ± 6 bar (233 ± 81 psi) during balloon kyphoplasty (p = 0.014). There were no significant differences in radiation exposure time.None of the patients underwent revision surgery, and postoperative neurologic sequelae were not observed. Cement leakage occurred at twenty-five of the 100 vertebral levels without significant differences between the two intervention arms (p = 0.230). Intraoperative material-related complications were observed at one of the fifty vertebral levels in the balloon kyphoplasty group and at nine of the fifty levels in the vertebral body stenting group. CONCLUSIONS: No beneficial effect of vertebral body stenting over balloon kyphoplasty was found among patients with painful osteoporotic vertebral fractures with regard to kyphotic correction, cement leakage, radiation exposure time, or neurologic sequelae. Vertebral body stenting was associated with significantly higher pressures during balloon inflation and more material-related complications.

Abstract

BACKGROUND: In the treatment of vertebral compression fractures, vertebral body stenting with an expandable scaffold inserted before application of the bone cement was developed to impede secondary loss of vertebral height encountered in patients treated with balloon kyphoplasty. The purpose of this study was to clarify whether there are relevant differences between balloon kyphoplasty and vertebral body stenting with regard to perioperative and postoperative findings. METHODS: In a two-armed randomized controlled trial, patients with a total of 100 fresh osteoporotic vertebral compression fractures were treated with either balloon kyphoplasty or vertebral body stenting. The primary outcome was the post-interventional change in the kyphotic angle on radiographs. The secondary outcomes were the maximum pressure of the balloon tamp during inflation, radiation exposure time, perioperative complications, and cement leakage. RESULTS: The mean reduction (and standard deviation) of kyphosis (the kyphotic correction angle) was 4.5° ± 3.6° after balloon kyphoplasty and 4.7° ± 4.2° after vertebral body stenting (p = 0.972). The mean pressures were 24 ± 5 bar (348 ± 72 pounds per square inch [psi]) during vertebral body stenting and 16 ± 6 bar (233 ± 81 psi) during balloon kyphoplasty (p = 0.014). There were no significant differences in radiation exposure time.None of the patients underwent revision surgery, and postoperative neurologic sequelae were not observed. Cement leakage occurred at twenty-five of the 100 vertebral levels without significant differences between the two intervention arms (p = 0.230). Intraoperative material-related complications were observed at one of the fifty vertebral levels in the balloon kyphoplasty group and at nine of the fifty levels in the vertebral body stenting group. CONCLUSIONS: No beneficial effect of vertebral body stenting over balloon kyphoplasty was found among patients with painful osteoporotic vertebral fractures with regard to kyphotic correction, cement leakage, radiation exposure time, or neurologic sequelae. Vertebral body stenting was associated with significantly higher pressures during balloon inflation and more material-related complications.

Statistics

Citations

11 citations in Web of Science®
12 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 27 Jun 2013
33 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Trauma Surgery
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:April 2013
Deposited On:27 Jun 2013 06:09
Last Modified:05 Jan 2017 14:49
Publisher:Boston, Journal of Bone and Joint Surgery
ISSN:0021-9355
Additional Information:Copyright: Journal of Bone & Joint Surgery
Publisher DOI:https://doi.org/10.2106/JBJS.L.00024
PubMed ID:23553291

Download

Preview Icon on Download
Preview
Content: Published Version
Filetype: PDF
Size: 978kB
View at publisher