Header

UZH-Logo

Maintenance Infos

A new implantable middle ear hearing device for mixed hearing loss: A feasibility study in human temporal bones


Huber, A M; Ball, G R; Veraguth, D; Dillier, N; Bodmer, D; Sequeira, D (2006). A new implantable middle ear hearing device for mixed hearing loss: A feasibility study in human temporal bones. Otology and Neurotology, 27(8):1104-1109.

Abstract

HYPOTHESIS: To assess the feasibility of a new, active middle ear device in temporal bones (TB). BACKGROUND: This device is designed for patients with mixed hearing loss subsequent to chronic middle ear infection, surgery, or trauma. This Bell-Vibroplasty is built from a VIBRANT MED-EL Vibrant Soundbridge and a Kurz Bell titanium partial ossicular replacement prosthesis. METHODS: In three fresh TBs, healthy and reconstructed middle ears were analyzed by means of laser Doppler interferometry. The sound transmission properties of a partial ossicular replacement prosthesis and a passive and an active Bell-Vibroplasty were compared with healthy middle ear function. RESULTS: The measurements provided reliable results with small standard deviations and good signal-to-noise ratios. The performance levels of the partial ossicular replacement prosthesis and of the passive Bell-Vibroplasty were comparable with that of healthy middle ear function. The activated Bell-Vibroplasty provided linear function and a flat frequency response within the measured frequency range (500 Hz-8 kHz), with peak deviations of less than 10 dB. The maximum output of the Bell-Vibroplasty was equivalent to 125-dB sound pressure level. CONCLUSION: Bell-Vibroplasty is feasible in TBs. Bell-Vibroplasty performance in TBs is sufficient to allow for a clinical trial as a next step.

Abstract

HYPOTHESIS: To assess the feasibility of a new, active middle ear device in temporal bones (TB). BACKGROUND: This device is designed for patients with mixed hearing loss subsequent to chronic middle ear infection, surgery, or trauma. This Bell-Vibroplasty is built from a VIBRANT MED-EL Vibrant Soundbridge and a Kurz Bell titanium partial ossicular replacement prosthesis. METHODS: In three fresh TBs, healthy and reconstructed middle ears were analyzed by means of laser Doppler interferometry. The sound transmission properties of a partial ossicular replacement prosthesis and a passive and an active Bell-Vibroplasty were compared with healthy middle ear function. RESULTS: The measurements provided reliable results with small standard deviations and good signal-to-noise ratios. The performance levels of the partial ossicular replacement prosthesis and of the passive Bell-Vibroplasty were comparable with that of healthy middle ear function. The activated Bell-Vibroplasty provided linear function and a flat frequency response within the measured frequency range (500 Hz-8 kHz), with peak deviations of less than 10 dB. The maximum output of the Bell-Vibroplasty was equivalent to 125-dB sound pressure level. CONCLUSION: Bell-Vibroplasty is feasible in TBs. Bell-Vibroplasty performance in TBs is sufficient to allow for a clinical trial as a next step.

Statistics

Citations

Dimensions.ai Metrics
39 citations in Web of Science®
42 citations in Scopus®
66 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 30 Mar 2009
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2006
Deposited On:30 Mar 2009 10:56
Last Modified:19 Feb 2018 06:46
Publisher:Lippincott Wiliams & Wilkins
ISSN:1531-7129
OA Status:Closed
Publisher DOI:https://doi.org/10.1097/01.mao.0000244352.49824.e6
PubMed ID:17031322

Download