Impact of a new motion-correction algorithm on image quality of low-dose coronary CT angiography in patients with insufficient heart rate control


Fuchs, Tobias A; Stehli, Julia; Dougoud, Svetlana; Fiechter, Michael; Sah, Bert-Ram; Buechel, Ronny R; Bull, Sacha; Gaemperli, Oliver; Kaufmann, Philipp A (2014). Impact of a new motion-correction algorithm on image quality of low-dose coronary CT angiography in patients with insufficient heart rate control. Academic Radiology, 21(3):312-317.

Abstract

RATIONALE AND OBJECTIVES: Prospective electrocardiogram (ECG) triggering allows coronary computed tomography angiography (CCTA) scanning with low radiation dose but requires heart rates below 63 beats/min. We assessed the impact of a novel vendor-specific motion-correction algorithm on image quality and interpretability of low-dose CCTA acquired despite insufficient heart rate control.
MATERIALS AND METHODS: In 40 patients undergoing CCTA for the assessment of known or suspected coronary artery disease who did not reach the target heart rate below 63 beats/min despite β-blockade before prospective low-dose scanning, the temporal acquisition window was increased (80 ms additional padding). The new algorithm detects and integrates vessel path and velocity from adjacent cardiac phases for motion correction. Two blinded observers assessed image quality on a 4-point Likert scale (1, nonevaluative; 2, reduced but evaluative; 3, good; and 4, excellent) and the fraction of interpretable segments (score 2 or more) using motion correction versus standard reconstruction.
RESULTS: Image reconstruction with motion correction resulted in an increased median coronary artery image quality score (excellent interobserver agreement, κ = 0.85) compared to standard reconstruction (3.4 vs. 3.0, P < .001). Consequently, motion-corrected reconstruction significantly improved the overall interpretability of coronary arteries (from 78% to 88%, P < .001). Estimated mean effective radiation dose was 2.3 ± 0.8 mSv.
CONCLUSIONS: A novel, vendor-specific, motion-corrected, reconstruction algorithm improves image quality and interpretability of prospectively ECG-triggered low-dose CCTA despite insufficient heart rate control.

Abstract

RATIONALE AND OBJECTIVES: Prospective electrocardiogram (ECG) triggering allows coronary computed tomography angiography (CCTA) scanning with low radiation dose but requires heart rates below 63 beats/min. We assessed the impact of a novel vendor-specific motion-correction algorithm on image quality and interpretability of low-dose CCTA acquired despite insufficient heart rate control.
MATERIALS AND METHODS: In 40 patients undergoing CCTA for the assessment of known or suspected coronary artery disease who did not reach the target heart rate below 63 beats/min despite β-blockade before prospective low-dose scanning, the temporal acquisition window was increased (80 ms additional padding). The new algorithm detects and integrates vessel path and velocity from adjacent cardiac phases for motion correction. Two blinded observers assessed image quality on a 4-point Likert scale (1, nonevaluative; 2, reduced but evaluative; 3, good; and 4, excellent) and the fraction of interpretable segments (score 2 or more) using motion correction versus standard reconstruction.
RESULTS: Image reconstruction with motion correction resulted in an increased median coronary artery image quality score (excellent interobserver agreement, κ = 0.85) compared to standard reconstruction (3.4 vs. 3.0, P < .001). Consequently, motion-corrected reconstruction significantly improved the overall interpretability of coronary arteries (from 78% to 88%, P < .001). Estimated mean effective radiation dose was 2.3 ± 0.8 mSv.
CONCLUSIONS: A novel, vendor-specific, motion-corrected, reconstruction algorithm improves image quality and interpretability of prospectively ECG-triggered low-dose CCTA despite insufficient heart rate control.

Statistics

Citations

16 citations in Web of Science®
16 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2014
Deposited On:13 Jan 2014 16:14
Last Modified:05 Apr 2016 17:21
Publisher:Elsevier
ISSN:1076-6332
Publisher DOI:https://doi.org/10.1016/j.acra.2013.10.014
PubMed ID:24332603

Download

Full text not available from this repository.
View at publisher