Influence of different maintenance times of torque application on the removal torque values to loosen the prosthetic abutment screws of external hexagon implants


Sella, Gustavo Castellazzi; Lopes Pereira Neto, Armando Rodrigues; Maziero Volpato, Cláudia Angela; de Vasconcellos, Diego Klee; Pekkan, Gurel; Özcan, Mutlu (2013). Influence of different maintenance times of torque application on the removal torque values to loosen the prosthetic abutment screws of external hexagon implants. Implant Dentistry, 22(5):534-539.

Abstract

PURPOSE: To analyze the torque application on prosthetic abutment screws using different maintenance times, to determine an influence on the removal torque values.
MATERIALS AND METHODS: A total of 40 external hexagon implants, 40 titanium screws, and 40 customized abutments were used. In group 1, the screws received a torque of 30 N·cm by instant torque application; in groups 2, 3, and 4, torque of 30 N·cm was applied and maintained for 10, 20, and 30 seconds, respectively. Removal torque was performed 10 minutes after torque application. Data were statistically analyzed using 1-way ANOVA and Tukey HSD test (α = 0.05).
RESULTS: The mean and standard deviation (±SD) of removal torque values found were 11.61 ± 1.43 N·cm for group 1; 18.64 ± 1.71 N·cm for group 2; 21.62 ± 0.97 N·cm for group 3; and 21.48 ± 1.55 N·cm for group 4. Groups 3 and 4 exhibited statistically higher values than group 2, which demonstrated significantly higher values than group 1 (P < 0.05).
CONCLUSIONS: A torque of 30 N·cm applied for 20 seconds seemed to be the best option when considering the removal torque values of external hexagon implants.

Abstract

PURPOSE: To analyze the torque application on prosthetic abutment screws using different maintenance times, to determine an influence on the removal torque values.
MATERIALS AND METHODS: A total of 40 external hexagon implants, 40 titanium screws, and 40 customized abutments were used. In group 1, the screws received a torque of 30 N·cm by instant torque application; in groups 2, 3, and 4, torque of 30 N·cm was applied and maintained for 10, 20, and 30 seconds, respectively. Removal torque was performed 10 minutes after torque application. Data were statistically analyzed using 1-way ANOVA and Tukey HSD test (α = 0.05).
RESULTS: The mean and standard deviation (±SD) of removal torque values found were 11.61 ± 1.43 N·cm for group 1; 18.64 ± 1.71 N·cm for group 2; 21.62 ± 0.97 N·cm for group 3; and 21.48 ± 1.55 N·cm for group 4. Groups 3 and 4 exhibited statistically higher values than group 2, which demonstrated significantly higher values than group 1 (P < 0.05).
CONCLUSIONS: A torque of 30 N·cm applied for 20 seconds seemed to be the best option when considering the removal torque values of external hexagon implants.

Statistics

Citations

1 citation in Web of Science®
1 citation in Scopus®
Google Scholar™

Altmetrics

Downloads

101 downloads since deposited on 05 Feb 2014
45 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Fixed and Removable Prosthodontics
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:05 Feb 2014 11:41
Last Modified:21 Dec 2016 08:45
Publisher:Lippincott, Williams & Wilkins
ISSN:1056-6163
Additional Information:The accepted version is a non-final version of an article published in final form in Implant Dentistry: October 2013 - Volume 22 - Issue 5 - p 534-539
Publisher DOI:https://doi.org/10.1097/ID.0b013e31829e548d
PubMed ID:23985429

Download