Fingerhut, R; Torresani, T (2013). Determination of total galactose from dried blood spots—extensive assay evaluation of a CE-marked test-kit. Journal of Analytical Sciences, Methods and Instrumentation, 3(3):163-166.

Abstract

Most newborn screening laboratories use CE-marked or FDA-approved test-kits, like in routine clinical chemistry. National regulations require only minimal evaluation from the customer, if the test-kits are used as specified by the manufacturer. The microtiter-based kit-concept is often based on the perception, that the laboratory always processes whole microtiter plates. However, in the daily routine, this is rather a rare exception, which leads to much higher costs per newborn, compared to the costs per assay in the test-kits. In addition the amount of wasted resources is quite high. Performance of the Neonatal Total Galactose kit from Perkin Elmer was tested. We have determined specificity, limit of detection (LOD), limit of quantitation (LOQ), intra and inter assay variation, recovery, stability of measuring signal and reagents. Results were also compared with the Astoria Pacific Spot Check System. In addition, we had (by chance) the opportunity to test 2 kits, which were already expired for more than 3 years. LOD was 165 - 306 μmol/L and LOQ 475 - 703 μmol/L, depending on the definition of LOD/LOQ. Mean recovery was 112.8%, intra assay CVs were 11.3, 7.3, 4.0, and 3.0, and inter assay CVs 28.7, 15.9, 7.8, and 9.3, at 220, 590, 1200, and 2060 μmol/L respectively. Reconstituted and mixed reagents must be used within some hours, and were unstable even if stored at -20℃. However, if the reconstituted galactose substrate reagent and galactose oxidase reagent were only mixed according to the daily requirements, and the rest stored separately at -20℃, they were stable for at least 12 days. The performance of the expired test-kits did not differ from the others. The performance of the Total Galactose kit is comparable to other tests used for newborn screening. However, we could significantly reduce the costs per newborn and reduce unnecessary production of waste, by thorough validation and modification of the assay procedures.

Abstract

Most newborn screening laboratories use CE-marked or FDA-approved test-kits, like in routine clinical chemistry. National regulations require only minimal evaluation from the customer, if the test-kits are used as specified by the manufacturer. The microtiter-based kit-concept is often based on the perception, that the laboratory always processes whole microtiter plates. However, in the daily routine, this is rather a rare exception, which leads to much higher costs per newborn, compared to the costs per assay in the test-kits. In addition the amount of wasted resources is quite high. Performance of the Neonatal Total Galactose kit from Perkin Elmer was tested. We have determined specificity, limit of detection (LOD), limit of quantitation (LOQ), intra and inter assay variation, recovery, stability of measuring signal and reagents. Results were also compared with the Astoria Pacific Spot Check System. In addition, we had (by chance) the opportunity to test 2 kits, which were already expired for more than 3 years. LOD was 165 - 306 μmol/L and LOQ 475 - 703 μmol/L, depending on the definition of LOD/LOQ. Mean recovery was 112.8%, intra assay CVs were 11.3, 7.3, 4.0, and 3.0, and inter assay CVs 28.7, 15.9, 7.8, and 9.3, at 220, 590, 1200, and 2060 μmol/L respectively. Reconstituted and mixed reagents must be used within some hours, and were unstable even if stored at -20℃. However, if the reconstituted galactose substrate reagent and galactose oxidase reagent were only mixed according to the daily requirements, and the rest stored separately at -20℃, they were stable for at least 12 days. The performance of the expired test-kits did not differ from the others. The performance of the Total Galactose kit is comparable to other tests used for newborn screening. However, we could significantly reduce the costs per newborn and reduce unnecessary production of waste, by thorough validation and modification of the assay procedures.

Statistics

Altmetrics

Downloads

62 downloads since deposited on 10 Feb 2014
18 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, not refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Medical Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:10 Feb 2014 13:54
Last Modified:05 Apr 2016 17:30
Publisher:Scientific Research Publishing, Inc.
ISSN:2164-2745
Publisher DOI:https://doi.org/10.4236/jasmi.2013.33020

Download

Preview Icon on Download
Preview
Content: Published Version
Filetype: PDF
Size: 262kB
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations