Customized covered stent graft for percutaneous closure of Fontan baffle leak


Binder, Ronald K; Nietlispach, Fabian; Carere, Ronald G (2013). Customized covered stent graft for percutaneous closure of Fontan baffle leak. Journal of Invasive Cardiology, 25(5):E118-E122.

Abstract

We report the innovation of both a partly-covered and completely-covered, variable-diameter, balloon-expandable stent that was custom-designed by NuMed, Inc for percutaneous closure of a baffle leak after total caval pulmonary connection (TCPC). A 50-year-old patient, born with tricuspid atresia, who had undergone TCPC, developed severe persistent cyanosis due to a right-to-left shunt through a TCPC baffle leak. Re-operation was deemed too high risk. Therefore, considering his complex anatomy, a custom-made, partly-covered, tapered, balloon-expandable stent was designed and successfully deployed. Months later, the shunt recurred at the junction of the covered to uncovered stent cells; subsequently, a second fully-covered, custom-made, tapered stent was implanted with an excellent outcome. The use of covered or partly-covered customized variable-diameter stents for closure of baffle leak after TCPC is feasible and resulted in marked clinical improvement. Customized balloon-expandable stents may be superior to traditional cylindrical stents because of better anchoring and apposition.

Abstract

We report the innovation of both a partly-covered and completely-covered, variable-diameter, balloon-expandable stent that was custom-designed by NuMed, Inc for percutaneous closure of a baffle leak after total caval pulmonary connection (TCPC). A 50-year-old patient, born with tricuspid atresia, who had undergone TCPC, developed severe persistent cyanosis due to a right-to-left shunt through a TCPC baffle leak. Re-operation was deemed too high risk. Therefore, considering his complex anatomy, a custom-made, partly-covered, tapered, balloon-expandable stent was designed and successfully deployed. Months later, the shunt recurred at the junction of the covered to uncovered stent cells; subsequently, a second fully-covered, custom-made, tapered stent was implanted with an excellent outcome. The use of covered or partly-covered customized variable-diameter stents for closure of baffle leak after TCPC is feasible and resulted in marked clinical improvement. Customized balloon-expandable stents may be superior to traditional cylindrical stents because of better anchoring and apposition.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

26 downloads since deposited on 13 Feb 2014
12 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:2013
Deposited On:13 Feb 2014 14:56
Last Modified:16 Feb 2018 19:49
Publisher:H M P Communications, LLC
ISSN:1042-3931
OA Status:Green
Free access at:Official URL. An embargo period may apply.
Official URL:http://www.invasivecardiology.com/files/wm%20E118-E122%20May%202013%20Binder.pdf
PubMed ID:23645059

Download

Download PDF  'Customized covered stent graft for percutaneous closure of Fontan baffle leak'.
Preview
Content: Published Version
Filetype: PDF
Size: 1MB