Comparing eye movements recorded by search coil and infrared eye tracking


Schmitt, Kai-Uwe; Muser, Markus H; Lanz, Christian; Walz, Felix; Schwarz, Urs (2007). Comparing eye movements recorded by search coil and infrared eye tracking. Journal of Clinical Monitoring and Computing, 21(1):49-53.

Abstract

OBJECTIVE: The performance of a new video-based infrared eye tracker (IR) was compared to the magnetic search coil technique (SC). Since the IR offers interesting possibilities as a diagnostic tool in neuro-ophthalmology, it was investigated whether the new device has overcome shortcomings that were reported from former IR systems.
METHODS: Horizontal saccades were recorded using the IR and the SC. The IR allowed eye movement recordings at different sampling rates ranging from 250 Hz to 1000 Hz while the SC recorded at 1000 Hz.
RESULTS/CONCLUSIONS: The results show that the IR and the SC were in good agreement and produced similar results. In contrast to other studies, the influence of the sampling rate of the IR was small. The saccade main-sequences did not show significant differences. The latency times observed for both systems were mainly in the short-latency range.

Abstract

OBJECTIVE: The performance of a new video-based infrared eye tracker (IR) was compared to the magnetic search coil technique (SC). Since the IR offers interesting possibilities as a diagnostic tool in neuro-ophthalmology, it was investigated whether the new device has overcome shortcomings that were reported from former IR systems.
METHODS: Horizontal saccades were recorded using the IR and the SC. The IR allowed eye movement recordings at different sampling rates ranging from 250 Hz to 1000 Hz while the SC recorded at 1000 Hz.
RESULTS/CONCLUSIONS: The results show that the IR and the SC were in good agreement and produced similar results. In contrast to other studies, the influence of the sampling rate of the IR was small. The saccade main-sequences did not show significant differences. The latency times observed for both systems were mainly in the short-latency range.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Language:English
Date:2007
Deposited On:28 Apr 2014 09:21
Last Modified:18 Feb 2018 13:13
Publisher:Springer
ISSN:1387-1307
OA Status:Closed
Publisher DOI:https://doi.org/10.1007/s10877-006-9057-5
PubMed ID:17120108

Download

Full text not available from this repository.
View at publisher

Get full-text in a library