Monitoring, analysis, and correction of magnetic field fluctuations in echo planar imaging time series


Kasper, Lars; Bollmann, Saskia; Vannesjo, S Johanna; Gross, Simon; Haeberlin, Maximilian; Dietrich, Benjamin E; Pruessmann, Klaas P (2015). Monitoring, analysis, and correction of magnetic field fluctuations in echo planar imaging time series. Magnetic Resonance in Medicine, 74(2):396-409.

Abstract

Purpose: To assess the utility of concurrent magnetic field monitoring for observing and correcting for variations in k-space trajectories and global background fields that occur in single-shot echo planar imaging (EPI) time series as typically used in functional MRI (fMRI).
Methods: Field monitoring was performed using an array of NMR field probes operated concurrently with series of single-shot EPI acquisitions from a static phantom. The observed fluctuations in field evolution were analyzed in terms of their temporal and spatial behavior at the field level as well as at the level of reconstructed image series. The potential to correct for such fluctuations was assessed by accounting for them upon image reconstruction. An indication of the number and relative magnitude of underlying effects was obtained via principal component analysis.
Results: Trajectory and global field variations were found to induce substantial image fluctuations. Global field fluctuations induced standard deviations in image intensity up to 31%. Fluctuations in the trajectory induced ghosting artifacts with standard deviations up to 2%. Concurrent magnetic field monitoring reduced the fluctuations in the EPI time series to a maximum of 1.2%.
Conclusion: Concurrent magnetic field monitoring holds the potential to improve the net sensitivity of fMRI by reducing signal fluctuations unrelated to brain activity

Abstract

Purpose: To assess the utility of concurrent magnetic field monitoring for observing and correcting for variations in k-space trajectories and global background fields that occur in single-shot echo planar imaging (EPI) time series as typically used in functional MRI (fMRI).
Methods: Field monitoring was performed using an array of NMR field probes operated concurrently with series of single-shot EPI acquisitions from a static phantom. The observed fluctuations in field evolution were analyzed in terms of their temporal and spatial behavior at the field level as well as at the level of reconstructed image series. The potential to correct for such fluctuations was assessed by accounting for them upon image reconstruction. An indication of the number and relative magnitude of underlying effects was obtained via principal component analysis.
Results: Trajectory and global field variations were found to induce substantial image fluctuations. Global field fluctuations induced standard deviations in image intensity up to 31%. Fluctuations in the trajectory induced ghosting artifacts with standard deviations up to 2%. Concurrent magnetic field monitoring reduced the fluctuations in the EPI time series to a maximum of 1.2%.
Conclusion: Concurrent magnetic field monitoring holds the potential to improve the net sensitivity of fMRI by reducing signal fluctuations unrelated to brain activity

Statistics

Citations

Dimensions.ai Metrics
12 citations in Web of Science®
12 citations in Scopus®
15 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 15 Sep 2014
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Uncontrolled Keywords:Radiology Nuclear Medicine and imaging
Language:English
Date:2015
Deposited On:15 Sep 2014 13:31
Last Modified:18 Aug 2018 16:53
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:0740-3194
OA Status:Closed
Publisher DOI:https://doi.org/10.1002/mrm.25407
PubMed ID:25123595

Download