Interdependencies of acquisition, detection, and reconstruction techniques on the accuracy of iodine quantification in varying patient sizes employing dual-energy CT.


Marin, Daniele; Pratts-Emanuelli, Jose J; Mileto, Achille; Husarik, Daniela B; Bashir, Mustafa R; Nelson, Rendon C; Boll, Daniel T (2015). Interdependencies of acquisition, detection, and reconstruction techniques on the accuracy of iodine quantification in varying patient sizes employing dual-energy CT. European Radiology, 25(3):679-686.

Abstract

PURPOSE To assess the impact of patient habitus, acquisition parameters, detector efficiencies, and reconstruction techniques on the accuracy of iodine quantification using dual-source dual-energy CT (DECT). MATERIALS AND METHODS Two phantoms simulating small and large patients contained 20 iodine solutions mimicking vascular and parenchymal enhancement from saline isodensity to 400 HU and 30 iodine solutions simulating enhancement of the urinary collecting system from 400 to 2,000 HU. DECT acquisition (80/140 kVp and 100/140 kVp) was performed using two DECT systems equipped with standard and integrated electronics detector technologies. DECT raw datasets were reconstructed using filtered backprojection (FBP), and iterative reconstruction (SAFIRE I/V). RESULTS Accuracy for iodine quantification was significantly higher for the small compared to the large phantoms (9.2 % ± 7.5 vs. 24.3 % ± 26.1, P = 0.0001), the integrated compared to the conventional detectors (14.8 % ± 20.6 vs. 18.8 % ± 20.4, respectively; P = 0.006), and SAFIRE V compared to SAFIRE I and FBP reconstructions (15.2 % ± 18.1 vs. 16.1 % ± 17.6 and 18.9 % ± 20.4, respectively; P ≤ 0.003). A significant synergism was observed when the most effective detector and reconstruction techniques were combined with habitus-adapted dual-energy pairs. CONCLUSION In a second-generation dual-source DECT system, the accuracy of iodine quantification can be substantially improved by an optimal choice and combination of acquisition parameters, detector, and reconstruction techniques. KEY POINTS • Iodine quantification techniques are not immune to error • Systematic deviations between the measured and true iodine concentrations exist • Acquisition parameters, detector efficiencies, and reconstruction techniques impact accuracy of iodine quantification.

Abstract

PURPOSE To assess the impact of patient habitus, acquisition parameters, detector efficiencies, and reconstruction techniques on the accuracy of iodine quantification using dual-source dual-energy CT (DECT). MATERIALS AND METHODS Two phantoms simulating small and large patients contained 20 iodine solutions mimicking vascular and parenchymal enhancement from saline isodensity to 400 HU and 30 iodine solutions simulating enhancement of the urinary collecting system from 400 to 2,000 HU. DECT acquisition (80/140 kVp and 100/140 kVp) was performed using two DECT systems equipped with standard and integrated electronics detector technologies. DECT raw datasets were reconstructed using filtered backprojection (FBP), and iterative reconstruction (SAFIRE I/V). RESULTS Accuracy for iodine quantification was significantly higher for the small compared to the large phantoms (9.2 % ± 7.5 vs. 24.3 % ± 26.1, P = 0.0001), the integrated compared to the conventional detectors (14.8 % ± 20.6 vs. 18.8 % ± 20.4, respectively; P = 0.006), and SAFIRE V compared to SAFIRE I and FBP reconstructions (15.2 % ± 18.1 vs. 16.1 % ± 17.6 and 18.9 % ± 20.4, respectively; P ≤ 0.003). A significant synergism was observed when the most effective detector and reconstruction techniques were combined with habitus-adapted dual-energy pairs. CONCLUSION In a second-generation dual-source DECT system, the accuracy of iodine quantification can be substantially improved by an optimal choice and combination of acquisition parameters, detector, and reconstruction techniques. KEY POINTS • Iodine quantification techniques are not immune to error • Systematic deviations between the measured and true iodine concentrations exist • Acquisition parameters, detector efficiencies, and reconstruction techniques impact accuracy of iodine quantification.

Statistics

Citations

5 citations in Web of Science®
5 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Date:March 2015
Deposited On:21 Oct 2014 12:38
Last Modified:05 Apr 2016 18:26
Publisher:Springer
ISSN:0938-7994
Publisher DOI:https://doi.org/10.1007/s00330-014-3447-8
PubMed ID:25278247

Download

Full text not available from this repository.
View at publisher

TrendTerms

TrendTerms displays relevant terms of the abstract of this publication and related documents on a map. The terms and their relations were extracted from ZORA using word statistics. Their timelines are taken from ZORA as well. The bubble size of a term is proportional to the number of documents where the term occurs. Red, orange, yellow and green colors are used for terms that occur in the current document; red indicates high interlinkedness of a term with other terms, orange, yellow and green decreasing interlinkedness. Blue is used for terms that have a relation with the terms in this document, but occur in other documents.
You can navigate and zoom the map. Mouse-hovering a term displays its timeline, clicking it yields the associated documents.

Author Collaborations