Publication: Massive aspiration past the tracheal tube cuff caused by closed trachael suction system
Massive aspiration past the tracheal tube cuff caused by closed trachael suction system
Date
Date
Date
Citations
Dave, M. H., Frotzler, A., Madjdpour, C., Koepfer, N., & Weiss, M. (2011). Massive aspiration past the tracheal tube cuff caused by closed trachael suction system. Journal of Intensive Care Medicine, 26(5), 326–329. https://doi.org/10.1177/0885066610392516
Abstract
Abstract
Abstract
Background: Aspiration past the tracheal tube cuff has been recognized to be a risk factor for the development of ventilator-associated pneumonia (VAP). This study investigated the effect of closed tracheal suctioning on aspiration of fluid past the tracheal tube cuff in an in vitro benchtop model. Methods: High-volume low pressure tube cuffs of 7.5 mm internal diameter (ID) were placed in a 22 mm ID artificial trachea connected to a test lung. Positive pressure ventilation (PPV) with 15 cm H(2)O peak inspiratory pressure and 5 cm H(2
Metrics
Views
Additional indexing
Creators (Authors)
Volume
Volume
Volume
Number
Number
Number
Page range/Item number
Page range/Item number
Page range/Item number
Page end
Page end
Page end
Item Type
Item Type
Item Type
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Language
Language
Language
Publication date
Publication date
Publication date
Date available
Date available
Date available
ISSN or e-ISSN
ISSN or e-ISSN
ISSN or e-ISSN
OA Status
OA Status
OA Status
Publisher DOI
Metrics
Views
Citations
Dave, M. H., Frotzler, A., Madjdpour, C., Koepfer, N., & Weiss, M. (2011). Massive aspiration past the tracheal tube cuff caused by closed trachael suction system. Journal of Intensive Care Medicine, 26(5), 326–329. https://doi.org/10.1177/0885066610392516