The Reamer-Irrigator-Aspirator (RIA) device represents a safe and efficient method to harvest autologous bone for grafting. However, hardware failure may occur, for example by breakage of the reamer head with metal debris remaining in the intramedullary canal. This case report describes the uncomplicated secondary removal of femoral intramedullary metal debris from a broken RIA reamer head; three weeks after the final surgery of a two-stage Masquelet procedure for the treatment of posttraumatic segmental bone loss at the tibia.
Additional indexing
Creators (Authors)
Volume
Volume
Volume
Page range/Item number
Page range/Item number
Page range/Item number
Item Type
Item Type
Item Type
In collections
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Language
Language
Language
Publication date
Publication date
Publication date
Date available
Date available
Date available
ISSN or e-ISSN
ISSN or e-ISSN
ISSN or e-ISSN
OA Status
OA Status
OA Status
Free Access at
Free Access at
Free Access at
Publisher DOI
Citations
Vetter, P., Hübner, C., Heining, S.-M., Hierholzer, C., & Pape, H.-C. (2024). Secondary removal of intramedullary metal debris from a defective Reamer-Irrigator-Aspirator (RIA) reamer head: A case report. Trauma Case Reports, 54, 101112. https://doi.org/10.1016/j.tcr.2024.101112