Publication: Impact of oxygen sources on performance of the Ventrain(®) ventilation device in an in vitro set-up
Impact of oxygen sources on performance of the Ventrain(®) ventilation device in an in vitro set-up
Date
Date
Date
Citations
Schmidt, A. R., Ruetzler, K., Haas, T., Schmitz, A., & Weiss, M. (2016). Impact of oxygen sources on performance of the Ventrain(®) ventilation device in an in vitro set-up. Acta Anaesthesiologica Scandinavica, 60, 241–249. https://doi.org/10.1111/aas.12663
Abstract
Abstract
Abstract
BACKGROUND The Ventrain(®) (Dolphys Medical, Eindhoven, The Netherlands) is a disposable handheld ventilation device allowing active inspiration and expiration through a transtracheal catheter. This study investigated Ventrain(®) 's performance when used with different clinical oxygen sources in an in vitro set-up. METHODS Three anesthesia oxygen sources (wall-mounted flowmeter, respirator oxygen outlet port, and anesthesia ventilator circuit) were used at gas flow rates of 6, 9, 12, and 15 l/min. First, the sources' driving pressu
Metrics
Views
Additional indexing
Creators (Authors)
Volume
Volume
Volume
Number
Number
Number
Page range/Item number
Page range/Item number
Page range/Item number
Page end
Page end
Page end
Item Type
Item Type
Item Type
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Language
Language
Language
Publication date
Publication date
Publication date
Date available
Date available
Date available
ISSN or e-ISSN
ISSN or e-ISSN
ISSN or e-ISSN
OA Status
OA Status
OA Status
Publisher DOI
Metrics
Views
Citations
Schmidt, A. R., Ruetzler, K., Haas, T., Schmitz, A., & Weiss, M. (2016). Impact of oxygen sources on performance of the Ventrain(®) ventilation device in an in vitro set-up. Acta Anaesthesiologica Scandinavica, 60, 241–249. https://doi.org/10.1111/aas.12663