Publication:

Noise figure characterization of preamplifiers at NMR frequencies

Date

Date

Date
2011
Journal Article
Published version

Citations

Citation copied

Nordmeyer-Massner, J. A., De Zanche, N., & Pruessmann, K. P. (2011). Noise figure characterization of preamplifiers at NMR frequencies. Journal of Magnetic Resonance Imaging (JMRI), 210(1), 7–15. https://doi.org/10.1016/j.jmr.2011.01.025

Abstract

Abstract

Abstract

A method for characterizing the noise figure of preamplifiers at NMR frequencies is presented. The noise figure of preamplifiers as used for NMR and MRI detection varies with source impedance and with the operating frequency. Therefore, to characterize a preamplifier's noise behavior, it is necessary to perform noise measurements at the targeted frequency while varying the source impedance with high accuracy. At high radiofrequencies, such impedance variation is typically achieved with transmission-line tuners, which however are not a

Additional indexing

Creators (Authors)

  • Nordmeyer-Massner, J A
    affiliation.icon.alt
  • De Zanche, N
    affiliation.icon.alt
  • Pruessmann, K P
    affiliation.icon.alt

Journal/Series Title

Journal/Series Title

Journal/Series Title

Volume

Volume

Volume
210

Number

Number

Number
1

Page range/Item number

Page range/Item number

Page range/Item number
7

Page end

Page end

Page end
15

Item Type

Item Type

Item Type
Journal Article

Dewey Decimal Classifikation

Dewey Decimal Classifikation

Dewey Decimal Classifikation

Language

Language

Language
English

Publication date

Publication date

Publication date
2011

Date available

Date available

Date available
2012-01-22

Publisher

Publisher

Publisher

ISSN or e-ISSN

ISSN or e-ISSN

ISSN or e-ISSN
1053-1807

OA Status

OA Status

OA Status
Closed

PubMed ID

PubMed ID

PubMed ID

Citations

Citation copied

Nordmeyer-Massner, J. A., De Zanche, N., & Pruessmann, K. P. (2011). Noise figure characterization of preamplifiers at NMR frequencies. Journal of Magnetic Resonance Imaging (JMRI), 210(1), 7–15. https://doi.org/10.1016/j.jmr.2011.01.025

Closed
Loading...
Thumbnail Image

Files

Files

Files
Files available to download:1

Files

Files

Files
Files available to download:1
Loading...
Thumbnail Image