dc.description.abstract

The present study aimed to compare the diagnostic reliability of a pre-set, manufacturer-specific, low-dose mode against a standard-dose mode in the detection of four different osseous lesions in the mandible with cone-beam computed tomography (CBCT). Four types of lesions (periapical lesion, extended periodontal gap, recession of the buccal lamella, sequestrum/fracture) were prepared on 40 pig mandibles. CBCT images were obtained from each mandible, with both the low-dose and standard-dose modes using the Orthophos SL CBCT device (Dentsply-Sirona, Bensheim, Germany). Twelve assessors quantitatively (detection of lesions) and qualitatively (assessment of detectability) evaluated the CBCT images in SIDEXIS 4 (Dentsply-Sirona) using a study-specific digital examination tool. A correct diagnosis was achieved in almost 71% (LD: 70.8%; SD: 70.9%) of 1920 lesions, without a statistically significant difference between the low-dose and standard-dose mode. This finding was consistent across all four lesion types. In conclusion, while low-dose mode and standard-dose mode CBCT scans performed similarly in the detection of four prepared lesions of the mandible, the former may be a promising, user-friendly alternative method of obtaining radiation-optimized, three-dimensional images in accordance with the As Low As Diagnostically Acceptable (ALADA) principle.

dc.identifier.doi10.3390/app10144961
dc.identifier.issn2076-3417
dc.identifier.scopus2-s2.0-85088652117
dc.identifier.urihttps://www.zora.uzh.ch/handle/20.500.14742/174896
dc.identifier.wos000557230600001
dc.language.isoeng
dc.subject.ddc610 Medicine & health
dc.title

Detectability of Osseous Lesions with a Pre-Programmed Low-Dose Protocol for Cone-Beam Computed Tomography

dc.typearticle
dcterms.accessRightsinfo:eu-repo/semantics/openAccess
dcterms.bibliographicCitation.journaltitleApplied Sciences
dcterms.bibliographicCitation.number14
dcterms.bibliographicCitation.originalpublishernameMDPI Publishing
dcterms.bibliographicCitation.pagestart4961
dcterms.bibliographicCitation.volume10
dspace.entity.typePublicationen
uzh.contributor.affiliationUniversity of Zurich
uzh.contributor.affiliationUniversity of Zurich
uzh.contributor.affiliationUniversity of Zurich
uzh.contributor.affiliationUniversity of Zurich
uzh.contributor.affiliationUniversity of Zurich
uzh.contributor.affiliationUniversity of Zurich
uzh.contributor.authorDöbelin, Quirin
uzh.contributor.authorStadlinger, Bernd
uzh.contributor.authorWiedemeier, Daniel B
uzh.contributor.authorBichsel, Dominique
uzh.contributor.authorRücker, Martin
uzh.contributor.authorValdec, Silvio
uzh.contributor.correspondenceYes
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.document.availabilitypublished_version
uzh.eprint.datestamp2020-12-04 13:03:55
uzh.eprint.lastmod2025-07-23 02:08:07
uzh.eprint.statusChange2020-12-04 13:03:55
uzh.harvester.ethYes
uzh.harvester.nbNo
uzh.identifier.doi10.5167/uzh-193064
uzh.jdb.eprintsId42538
uzh.oastatus.unpaywallgold
uzh.oastatus.zoraGold
uzh.publication.citationDöbelin, Quirin; Stadlinger, Bernd; Wiedemeier, Daniel B; Bichsel, Dominique; Rücker, Martin; Valdec, Silvio (2020). Detectability of Osseous Lesions with a Pre-Programmed Low-Dose Protocol for Cone-Beam Computed Tomography. Applied Sciences, 10(14):4961.
uzh.publication.freeAccessAtdoi
uzh.publication.originalworkoriginal
uzh.publication.publishedStatusfinal
uzh.scopus.impact6
uzh.scopus.subjectsGeneral Materials Science
uzh.scopus.subjectsInstrumentation
uzh.scopus.subjectsGeneral Engineering
uzh.scopus.subjectsProcess Chemistry and Technology
uzh.scopus.subjectsComputer Science Applications
uzh.scopus.subjectsFluid Flow and Transfer Processes
uzh.workflow.doajuzh.workflow.doaj.true
uzh.workflow.eprintid193064
uzh.workflow.fulltextStatuspublic
uzh.workflow.revisions50
uzh.workflow.rightsCheckkeininfo
uzh.workflow.sourceCrossRef:10.3390/app10144961
uzh.workflow.statusarchive
uzh.wos.impact5
Files

Original bundle

Name:
2020_Dobelin_et_al_DetectabilityOsseousLesionsLowDoseCBCT_ApplSci.pdf
Size:
795.32 KB
Format:
Adobe Portable Document Format
Publication available in collections: